![MedeScan 2 IN 1 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/medescan/2-in-1/2-in-1_manual_1759993023.webp)
Guidance and manufacturer´s declaration –
electromagnetic immunity –for EQUIPMENT and
SYSTEM that are not LIFE SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the Infrared
Thermometer should assure that it is used in such an environment.
Electromagnetic environment -
guidance
Immunity
test
IEC
60601
Test level
Compliance
Level
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz
to 2.5
GHz
10V/m
Portable and mobile RF commun-
ications equipment should be used
no closer to any part of the Infrared
Thermometer, including cables,than
the recommended separation dist-
ance calculated from the equation
applicable to the frequency of the
transmitter.Recommended
separation distance
Where p is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is rec-
ommended separation distance
in meters (m).
b
Field strengths from fixed RF tra-
nsmitters, as determined by an
Electromagnetic site survey, a
should be less than the complian-
ce level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
19
800 MHz to 2,5 GHz
80 MHz to 800 MHz
b