11
EN
Tubing care
Inspect the tubing after each pumping session for condensation
and/or milk.
If condensation appears in the tubing
Let the pump run after you have completed pumping with tubing
attached for 1– 2 minutes or until dry.
If milk gets into the tubing
If milk accidentally gets into the tubing turn off the pump and unplug
from power source. Remove and clean the tubing according to the
instructions «Prior to first use and once daily». Shake water droplets
out and hang to air-dry. For faster drying, attach the tubing to pump
and run for 1– 2 minutes or until dry.
Hints
After cleaning, when parts are not in use, put in a clean zip lock
bag or store in a container with lid. Alternatively, the parts can be
wrapped in a clean paper or cloth towel.
Cleaning
Disassemble and thoroughly wash all parts in warm, soapy water.
Rinse in clear water.
Disinfecting
Disinfect all parts by washing in a thermic washing machine
(at approx. 93° C).
Sterilizing
Prior sterilizing visually check parts if clean and intact. Sterilize all
parts by autoclaving at 134° C for 3 (or 18) minutes or at 125° C
for 15 – 20 minutes. Do not stack parts or apply external pressure
to avoid damages. Alternatively, the parts can be EO-sterilized
(Ethylene Oxide).
Hygiene
• Wash hands thoroughly with soap and water before touching
the breast and pump parts.
• Avoid touching the inside of bottles and lids.
• Clean the breast and surrounding area with a damp cloth.
• Do not use soap or alcohol which can dry the skin and cause
cracked nipples.