22 |
Battery data and test
11 Battery data and test
(AC / DC version only)
CAUTIONS
Only charge the rechargeable battery when the charging indicator blinks slowly�
The rechargeable battery is charged during mains operation�
Duration of pumping
If the battery is fully charged, the operating time at maximum vacuum is approximately
> 50 minutes�
Charging time
(charged up to 100 %)
Approx� 5 hours�
Pull out the mains plug from the fixed mains socket�
2�
Testing the
rechargeable battery
5� Measure the time until the
pump comes to a stop�
The battery is ok, when
the the running time is
> 50 minutes�
If the running time is
< 50 minutes, repeat test
or calibrate the battery
(see next page).
4�
4�1 Switch on the Clario� The
pump will start running�
3�
3�1 Set vacuum regulator to
position «max�»�
Make sure that the rechargeable battery is fully charged�
1�
not lit
Battery is fully charged
lit up
blinks slowly
Battery is soon empty
blinks rapidly
Battery is defective
– Battery is being charged
– Pump is plugged in
Battery status
Содержание Clario
Страница 330: ...Notes ...
Страница 331: ...Notes ...