Calibration Procedure
NOTE:
This procedure should be performed in the same
geographic area as the measurements for infant
weights. Procedure requires
two
certified 10 kg
Class F test weights.
1.
Press
button while the scale is off
2.
Press
1
button while the display
is flashing all segments --> Display
shows “Func 1”
3.
Press
reweigh
fed baby
button --> Display
shows “Func 2”
4.
Press
3
button --> Display
shows “9XXX” (For example, 9802
will be displayed)
NOTE:
The gravity setting will update automatically upon
completing the calibration procedure.