Alarms and Troubleshooting
43
9
Alarms and Troubleshooting
Warnings:
During troubleshooting by service personnel, there is a 100-240V AC high voltage in the HP-60 and
the switching power supply circuit may generate a 400V DC high voltage. An accidental touch may
cause serious personal injury. Please follow the requirements in related regulations and ensure that
the maintenance is performed by professional qualified personnel.
When a device is used to measure a circuit board or other parts, short circuit or incorrect
measurement method may cause serious personal injury or serious device damage.
The device safety and functions must be verified after being disassembled for maintenance and then
assembled each time. For details, see chapter 7.
Basic tools for maintenance and inspection:
PC (used for downloading software and filling table)
Software downloading connector
Digital multimeter
20-50 ml measuring cylinder (accuracy: 0.2 ml) or electronic balance
Time keeping device like a stopwatch
Notes:
Among the aforementioned tools, the measuring cylinder or electronic balance is used to check the
infusion accuracy.
For the tools required for disassembling and assembling the device, see chapter 8.
Dedicated safety test devices must be used to test safety items after disassembling and then
assembling the device.
9.1
Alarm Levels
The infusion pump provides users with a variety of status information about itself and its infusion process. If
any exception is detected, the infusion pump generates an alarm and informs users in the form of sound, light,
and character.
Based on the criticality of the exception information, alarms are classified into three levels: low, middle,
and high. The following table describes the sound-light presentation mode of these three levels of alarm. The
alarm volume ranges from 45 dB to 85 dB.
Alarm Level
Alarm Sound
Alarm Indicator
Status
Low
“Di” is given one time.
The
yellow
indicator is steady
on.
Middle
“Di-di-di” is given every 12s.
The
yellow
indicator flashes.