Sensor Information
25
5
Sensor Information
This chapter only describes the software UIs related to servicing and test. For the description of other
software functions, see the corresponding operation manual.
5.1
Sensor Information Test
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, and choose
Sensor Info
.
a)
Motor status test: Select
Clockwise
or
Anticlockwise
for the motor, select the start speed of the motor
(motor speed setting range: 0.01-150r/s), and check whether you can hear the motor rotation sound. If
you does not head the rotation sound, the motor or main control board may be faulty.
b)
Upper/lower pipeline check: When the upper/lower pipeline is equipped with an infusion tube, the
status displayed is
1
. Otherwise, the status displayed is
0
. If the display status is incorrect, the
optocoupler corresponding to the upper/lower pipeline may be damaged.
c)
Pump finger signal: When the pump finger board optocoupler is shielded, the status displayed is
1
.
Otherwise, the status displayed is
0
. If the display status is incorrect, check the pump finger
optocoupler and related connection cable.
d)
Air bubble detection: When no tube is installed or the tube is empty, a value in the range of 0-100 is
displayed. When a full tube is installed, a value larger than 1000 is displayed. Otherwise, check
whether the tube is installed in position and whether the air bubble sensor and detection board are
faulty.
e)
Dedicated consumable: Check whether the consumable used is a dedicated consumable. If yes, the
status displayed is
1
. Otherwise, the status displayed is
0
.
f)
AC test: Check whether a value in the range of 14V-16V is displayed when the pump is connected to
AC power supply and a value smaller than 2V is displayed when the pump is not connected to AC
power supply. If the displayed value is incorrect, the device generates the ERR007 alarm, which may
be caused by a power module failure.
Содержание HP-60
Страница 30: ...Maintenance 30 Please follow local laws to dispose the old battery...
Страница 35: ...Calibration 35 Schematic diagram...