6.  BONE GRAFT SHAPING

The bone graft should be shaped according to the planned/

desired dimensions.

Select the Graft Shaper according to the desired angulation 

and assemble it with the Graft Holder. Four different angles 

are available: 0°, 10°, 20° and 30°.

10.

Put the bone graft into the assembled instrument making 

sure to place the curved surface of the graft on the bottom 

of the device.

Rotate the knob to adjust the graft height according to the 

desired dimensions. The numbers on the Graft Shaper 

indicate the bone graft thickness. Depending on the chosen 

angle, the numbers indicate the thickness of the two sides 

of the graft.

Then, press the lateral pincer in order to firmly hold the graft 

in place. Use a HEX 3.5 screwdriver to tighten the screw 

inside the pincer to lock it.

11.

Perform the final shaping using an oscillating saw

12.

CAUTION

 

 

Pay attention to keep the hands away from the cutting 

plane while performing the resection.

Part numbers subject to change.

NOTE FOR STERILIZATION

The instrumentation is not sterile upon delivery. Instruments must be cleaned before use and sterilized in an autoclave respecting 

the US regulations, directives where applicable, and following the manufactures instructions for use of the autoclave.  

For detailed instructions please refer to the document “Recommendations for cleaning decontamination and sterilisation of 

Medacta International orthopaedic devices” available at www.medacta.com.

Содержание Shoulder System

Страница 1: ...Sports Med Joint Spine Surgical Technique COMPLETE CONVERTIBLE INNOVATIVE GRAB INSTRUMENTS...

Страница 2: ...GRAB Instruments Surgical Technique 2...

Страница 3: ...3 TABLE OF CONTENTS 1 INTRODUCTION 4 2 K WIRE POSITIONING 4 3 BACK SURFACE REAMING 4 4 BONE GRAFT REAMING 5 5 BONE GRAFT HARVESTING 6 6 BONE GRAFT SHAPING 7...

Страница 4: ...IRE POSITIONING Position the Humeral K wire Positioner onto the center of the humeral head in order to match the bone convexity Insert the 2 5mm K wire through the central hole of the Humeral K wire P...

Страница 5: ...he numbers on the external surface of the Graft Reamer represent the height of the bone graft If the Central Piston Guide is used as a mechanical stop the maximum bone reaming depth is 10mm or 20mm de...

Страница 6: ...t the resection plane is flush with the most medial insertion line of the supraspinatus and the shaft that follows the humeral diaphysis This will result in an approximate cut inclination of 135 Check...

Страница 7: ...f the graft Then press the lateral pincer in order to firmly hold the graft in place Use a HEX 3 5 screwdriver to tighten the screw inside the pincer to lock it 11 Perform the final shaping using an o...

Страница 8: ...ternational SA Strada Regina 6874 Castel San Pietro Switzerland Phone 41 91 696 60 60 Fax 41 91 696 60 66 info medacta ch Find your local dealer at medacta com locations All trademarks and registered...

Отзывы: