Maintenance
1.
The percussor cooling fan draws in 4 cubic feet per
minute of air. Therefore considerable dust and lint can be
drawn into the percussor filter.
CAUTION
– If the percussor filter becomes clogged
with lint deposits, the unit will shut off permanently.
2.
Inspect the percussion fan vent regularly. Use
compressed canned air once a month, or as needed, to
clean and remove lint from the fan located in the Power
Head. Spray air onto the fan opening, following usage
directions on the can.
CAUTION
– Do not insert extension tube into the
opening as is it might damage the fan.
Disposal
Please dispose of your retired percussor by contacting your
local landfill and requesting a designated e-waste drop off
location. Alternatively, you may return the retired unit to
Med Systems for disposal.