OPERATING INSTRUCTIONS

POSITIONING ELECTRODES

The 4 self-adhesive electrodes need to be placed correctly for maximum pain relief. For the
best results please ask for assistance from your birthing partners or medical professional.

Carefully remove 2 electrodes from the plastic backing and place approx. 2” (5cm) either
side of the spine just below the bra line as show in Fig8 and connect the patient lead from
channel 1 to these electrodes. Now position the bottom 2 electrodes as shown in Fig8
(middle panty line) and connect to channel 2 patient lead.

Bra line

Electrodes

Electrodes

In line with navel

Panty line

Fig8

MATERNITY MODE

Insert both patient

leads in to the sockets

on top of the device

Connecting the electrodes pads 

Insert the pins (at the end of each lead

wire) into each pair of electrodes

When connecting the

Patient leads, please

ensure they are inserted

completely straight and not

at an angle. Pushing at an

angle may bend the pins

Fig 6 

Fig7 

Содержание EV-806

Страница 1: ...ent Instruction User Manual Tel 0800 121 4626 email sales med fit co uk Med Fit TENS and Muscle Stimulator Maternity TENS Pelvic Floor Stimulator Muscle Toning Stimulator Pain Relief TENS Machine med...

Страница 2: ...x Mums to Be TENS 1 x SA20 pack of self adhesive electrode 1 x 9v PP3 battery not rechargeable 1 x Carry case 1 x Pair TENS patient leads 1 x User manual Total Tone Pelvic Floor Stimulator is supplie...

Страница 3: ...7 MATERNITY MODE OPERATING INSTRUCTIONS 8 PELVIC FLOOR STIMULATOR OPERATING INSTRUCTIONS 11 MUSCLE STIMULATOR OPERATING INSTRUCTIONS 16 TENS PAIN RELIEF OPERATING INSTRUCTIONS 20 GENERAL PRECAUTIONS W...

Страница 4: ...s unit is a combination stimulator of TENS and EMS which can be used for pain relief and muscle stimulation The device is provided with two controllable output channels each independent of the other A...

Страница 5: ...this mode Mode 2 Pelvic Floor Stimulation Programmes P3 to P6 are specially designed for the use in strengthing and conditioning the pelvic floor when used with a vaginal probe see pages 11 to 15 for...

Страница 6: ...ND FEATURES Boost Button Only operational with programmes P1 P2 Obstetic TENS mode Keylock facility 9v PP3 Battery Sockets for electrodes Channel 2 Intensity controls Channel 1 Intensity controls Mode...

Страница 7: ...ams Mode Prog Timer Pelvic floor toning P3 C Continuous Pelvic floor toning P4 C Continuous Pelvic floor toning P5 C Continuous Pelvic floor toning P6 C Continuous Pelvic Floor Toning Programs 7 www t...

Страница 8: ...2 each single press of this button changes the programme from P1 to P2 P3 etc to P12 If at any stage you require to go back to P1 please hold the mode button for approx 3 seconds and the device will r...

Страница 9: ...ine as show in Fig8 and connect the patient lead from channel 1 to these electrodes Now position the bottom 2 electrodes as shown in Fig8 middle panty line and connect to channel 2 patient lead Bra li...

Страница 10: ...ng a constant tingling sensation Once the contraction has passed simply switch the boost button back to its off position The TENS will return to the original burst settings Use this programme P1 for t...

Страница 11: ...S ARE SPECIFICALLY RECOMMENDED FOR THE FOLLOWING A Weak pelvic floor muscles B Bladder weakness C All urinary incontinence D Tighten tone and improve pelvic E Stress incontinence F Urge incontinence P...

Страница 12: ...the Intensity Decrease button to decrease the amplitude Once a programme is selected you have 60 seconds to adjust the amplitude to a comfortable level by pressing the On and Off buttons After the pro...

Страница 13: ...ually occurs because of problems with muscles that help to hold or release urine The kidneys excrete the urine which is stored in the bladder Urine exists from the bladder via the urethra Urination oc...

Страница 14: ...late nearby muscles This helps to stabilize overactive muscles and stimulate contraction of urethral muscles Electrical stimulation can be used to reduce both stress and urge incontinence The Stress I...

Страница 15: ...1 The incontinence stimulator should not be used during pregnancy 2 The device should be used only under the continued supervision of a physician 3 The user must keep the device out of the reach of ch...

Страница 16: ...uscle needing treatment this causes the muscle to contract and relax It is derived from the square wave from originally invented by John Faraday in 1831 It works by directly stimulating motor neurons...

Страница 17: ...lation Toning and Conditioning Programmes 7 8 9 10 Toning Abdominals Leg toning and massage POSITION OF ELECTRODES Best results are achieved by placing the electrodes directly over the pain site With...

Страница 18: ...Electrode Placement 18 Telephone 0800 121 4626 www tensmachineuk com Buttocks leg toning EMS STIMULATION MODE...

Страница 19: ...19 www tensmachineuk com Telephone 0800 121 4626 EMS STIMULATION MODE Electrode Placement Inner thigh toning Waist line toning shaping...

Страница 20: ...damage or injury to vital parts of our bodies Even though pain is a necessary warning signal of trauma or malfunction in the body nature may have gone too far in its design Aside from its value in dia...

Страница 21: ...ow high should I turn the intensity Everybody reacts differently to TENS Stimulation so it is important that you increase the intensity sensation feeling to the correct level Increase the intensity to...

Страница 22: ...grammes P11 P12 Primary Placement Shoulder Pain Suggested Programmes P11 P12 Primary Placement Alternative Placement Degenerative Arthritis Cervical and Lumbar Suggested Programmes P11 P12 Primary Pla...

Страница 23: ...ack Pain Suggested Programmes P11 P12 Primary Placement Hip Neuralgia Suggested Programmes P11 P12 Primary Placement Phantom Limb Lower Extremity Suggested Programmes P11 P12 Primary Placement Sciatic...

Страница 24: ...ment Primary Placement Wrist Pain Suggested Programmes P11 P12 Primary Placement Alternative Placement Carpal Tunnel Syndrome Suggested Programmes P11 P12 Primary Placement Alternative Placement Elbow...

Страница 25: ...Alternative Placement Primary Placement Knee Pain Post Op Suggested Programmes P11 P12 Primary Placement Alternative Placement Cervical Placement Suggested Programmes P11 P12 Primary Placement Alterna...

Страница 26: ...ectronic device should not undergo TENS treatment without consulting a medical professional Patients with a heart disease epilepsy cancer or any other health condition should not undergo TENS treatmen...

Страница 27: ...ge following trauma or fracture b Following recent surgical procedures when muscle contraction may disrupt the healing process c Over the menstruating or pregnant uterus and d Over areas of skin which...

Страница 28: ...rom 5 to 60 minutes and continue C 5 minutes each step Patient Compliance Meter This unit can store 60 sets of operation records Total recorded time is 999 hours Low Battery Indicator A low battery in...

Страница 29: ...insert the plug into AC power supply socket 3 Timer 4 Increment 5 Decrement 6 Consult instructions for use 7 Manufacturer 8 Serial number 9 Lock 10 Low battery 11 Pause 12 DC current DC Power source...

Страница 30: ...dust free environment until the gel regains its tack Removal and storage 1 Lift a corner of the electrode and slowly peel the electrode off the skin touching the adhesive gel as little as possible 2 P...

Страница 31: ...TENS EMS models carry a warranty of one year from the date of delivery The warranty applies to the stimulator only and covers both parts and labor relating thereto The warranty does not apply to damag...

Страница 32: ...IEC 61000 3 2 Voltage fluctuations Complies flicker emissions IEC 61000 3 3 This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC in...

Страница 33: ...ality should be IEC 61000 4 5 line s mode that of a typical commercial or hospital environment Voltage dips 5 UT 5 UT Mains power quality should be short 95 dip in UT 95 dip in UT that of a typical co...

Страница 34: ...xed RF transmitters as determined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with...

Страница 35: ...ording to frequency of transmitter output power of m transmitter W 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 68...

Страница 36: ...Tel 0161 427 4851 Fax 0161 427 0215 www tensmachineuk com...

Отзывы: