background image

FineTuner Echo

12

Additional functions of your FineTuner Echo

Testing your audio processor

You can use your FineTuner Echo to test proper function of the audio processor 
microphones: 

1.  Switch on the audio processor.
2.  Place the coil over the battery compartment of FineTuner Echo (see Fig. 5).
3.  When speaking into the microphone, the green light on the FineTuner Echo should 

flicker in the rhythm of your voice.

Fig. 5 Testing your audio processor

If the green light does not light up or stays on constantly, try the following:

1.  Adjust the volume setting. By using the appropriate loudness setting, you should be 

able to recognize the flickering of the green light in the rhythm of your voice.

2.  Change the batteries of your audio processor.
3.  Replace the existing coil cable with a substitute cable.

If these measures are not successful, contact your healthcare provider or local MED‑EL 
representative.

Monitoring mode

When pressing the MT button for 3 seconds, the FineTuner Echo changes into the 
monitoring mode. The symbol 

 appears on the bottom of the display. The monitoring 

mode is used with the SONNET MTD. For further information about this optional service 
tool for the MED‑EL Cochlear Implant System, please read the applicable user manual or 
ask your audiologist for assistance.

In addition, you can switch your audio processor to silent mode if required.

3

1

2

Содержание FineTuner Echo

Страница 1: ...Cochlear Implants User Manual FineTuner Echo AW4381_2 0 English...

Страница 2: ......

Страница 3: ...er Echo display 9 Additional functions of your FineTuner Echo 12 Wearing options for your FineTuner Echo 13 4 General precautions and warnings 14 5 Care and maintenance 16 Maintenance 16 Disposal 16 B...

Страница 4: ...teries The user of the FineTuner Echo does not need any special skills or elevated level of education however for young children and those who for a variety of reasons lack the ability to appropriatel...

Страница 5: ...nd also in a clinical environment e g audiologist s office Indications Any user having received a MED EL Cochlear Implant is a potential user of the FineTuner Echo Additionally any person that interac...

Страница 6: ...ily listening situations Its ergonomic design and display facilitate changing the settings of your audio processor The FineTuner Echo is not necessary for the function of your audio processor When swi...

Страница 7: ...FineTuner Echo 5 Fig 2 FineTuner Echo overview Display Keypad Battery lid Silicone ring Battery Battery compartment Green light Front view Back view...

Страница 8: ...coin cell battery type CR2032 with the sign facing up Hold the battery at a slight angle and insert it under the lip in the battery compartment Fig 3 2 then gently push it down 4 To close the lid ali...

Страница 9: ...other side Fig 4 Pairing the FineTuner Echo and the audio processor NOTE Your audio processor has to be configured for use with the FineTuner Echo If the audio processor is not programmed correctly th...

Страница 10: ...keypad Processor selection key Selects left audio processor Processor selection key Selects left and right audio processor Processor selection key Selects right audio processor Your FineTuner Echo do...

Страница 11: ...he filled square indicates the selected programme Microphone Sound is perceived via the audio processor microphones only Microphone and telecoil Sound is perceived via the audio processor microphones...

Страница 12: ...cho Pairing Screen The FineTuner Echo must be paired with an audio processor Pair your FineTuner Echo See chapter Using the FineTuner Echo for the first time for instructions Screen indicating success...

Страница 13: ...Interference from other electronic or electrical equipment is blocking the transmission Move the FineTuner Echo closer to your ear and try to switch settings again If you only move the FineTuner Echo...

Страница 14: ...dness setting you should be able to recognize the flickering of the green light in the rhythm of your voice 2 Change the batteries of your audio processor 3 Replace the existing coil cable with a subs...

Страница 15: ...end below the battery compartment To attach a lanyard thread it through the small openings as shown in Fig 6 Fig 6 Attaching the lanyard to the FineTuner Echo Do not use the lanyard with children unde...

Страница 16: ...between 5 C and 40 C for FineTuner Echo Avoid excessive heat Excessive heat might damage the FineTuner Echo and make it unusable If the FineTuner Echo should be at a temperature that is outside the d...

Страница 17: ...ipment including peripherals such as antenna cables and external antennas should be used no closer than 30cm to any part of the FineTuner Echo including cables specified by MED EL Otherwise degradatio...

Страница 18: ...lean the FineTuner Echo Never use household cleaning products or alcohol Do not use a microwave or any other heating devices to dry the FineTuner Echo If necessary wipe it off with a dry tissue Do not...

Страница 19: ...ir original packaging in a cool and dry place Isolate or cover the positive and negative terminals to avoid short circuiting Do not store nor use batteries in high temperature and high humidity locati...

Страница 20: ...new and used batteries out of the reach of children Children shall be instructed not to swallow or put any components of their MED EL Cochlear Implant System into their mouths or to play with any com...

Страница 21: ...ack silicone ring does not fall out If you lose the black silicone ring please order a replacement from MED EL 3 Replace the used coin cell battery by gently shaking it into your hand Try not to touch...

Страница 22: ...command The audio processor is switched off The battery of the audio processor is empty Interference from other electronic or electrical equipment is blocking the transmission Move the FineTuner Echo...

Страница 23: ...ut with the FineTuner Echo you need to switch the audio processor off and on again When the audio processor is switched on again it will automatically start with the microphone setting activated If yo...

Страница 24: ...wer supply 1 lithium manganese dioxide battery type CR2032 3V Internally powered Battery life expectancy is typically more than 6 months Materials Mixture of polycarbonate and acrylonitrile butadiene...

Страница 25: ...d service life Applied parts The surface of the FineTuner Echo which is in contact with the user is a Type BF Applied Part Degree of protection IP52 Dust protected Protected against vertically falling...

Страница 26: ...ul interference and 2 this device must accept any interference received including interference that may cause undesired operation Warning Changes or modifications made to this equipment not expressly...

Страница 27: ...ss www medel com compliance Type BF IEC 60601 1 EN 60601 1 The surface of the FineTuner Echo which is in contact with the user is a Type BF Applied Part Manufacturer Direct current Atmospheric pressur...

Страница 28: ...formation Non ionizing radiation Fragile handle with care Information indicating a hazardous situation that if not avoided could result in death or serious injury Information indicating a hazardous si...

Страница 29: ...etic environment guidance RF emissions CISPR 11 Group 1 The FineTuner Echo uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interfe...

Страница 30: ...Mains power quality should be that of a typical commercial or hospital environ ment 1kV for input output lines Surge IEC 61000 4 5 1kV line s to line s Not applicable Mains power quality should be th...

Страница 31: ...obile RF communi cations equipment should be used no closer than 30cm to any part of the FineTuner Echo including cables specified by MED EL Otherwise degradation of the performance of the FineTuner E...

Страница 32: ...l MED EL representative Warranty Please refer to the accompanying Warranty Statement for information on our warranty provisions Manufacturer address MED EL Elektromedizinische Ger te GmbH Worldwide He...

Страница 33: ...Contact MED EL 31 9 Contact MED EL Please refer to the accompanying Contact Sheet for your local office...

Страница 34: ...32...

Страница 35: ......

Страница 36: ...MED EL Elektromedizinische Ger te GmbH F rstenweg 77a 6020 Innsbruck Austria office medel com medel com...