13
trOUBlEshOOtiNg
7
TROUBLESHOOTING
7.1
Quick Troubleshooting Guide
If the equipment appears not to be working correctly, read the instructions again and then check
Problem
Possible Cause
Solution
an acoustic signal (3 beeps)
is produced during the
exposure cycle and at the
end of the cycle starlight s +
will not allow any further
treatment to be carried out.
the selected function lED
flashes.
the thermal protection has
been activated.
it will be possible to use
the equipment only after
it has cooled down. Wait
approximately 10 minutes
before re-using the device.
When the ‘start’ button
is pressed, an acoustic
warning sounds (4 beeps)
and the device does not
emit light.
the temperature control
circuit reports a failure.
Contact a Mectron service
centre.
When the ‘start’ button is
pressed, the mode selection
LED flashes. The device
does not emit light, or
the intensity of the light
emitted is unsuitable.
the control circuit reports an
lED failure.
Contact a Mectron service
centre.
the polymerisation is
insufficient.
the terminal surface of the
tip of the optical fibre is
soiled.
see
of
the Use and Maintenance
manual
.
Table 4 –
Quick troubleshooting guide
Содержание starlight s +