15
INSTALLATION AND CONFIGuRATION
Place the irrigation bag onto the
appropriate stand;
Remove the protection cap of the spike;
Insert the spike into the irrigation bag;
NOTA:
The irrigation bag holder stand
must be used with bags of up to a
maximum of 1000 ml capacity.
3
Open the air inlet on the spike, before
proceeding with the operation.
Open the irrigation tube clamps, if closed;
4
4.6
Connection of the Handpieces to the Device
Channel 1 (CH1) allows the connection of
the PIEZOSURGERY
®
medical handpiece
that can be equipped with the sterile,
single-use PIEZOSuRGERY
®
medical
inserts.
Channel 2 (CH2) allows the connection of
the PIEZOSURGERY
®
handpiece
that can be equipped with the sterile,
single-use PIEZOSuRGERY
®
inserts.
Insert the handpiece connector into
the suitable port located on the front
panel of the device, assuring that the
index point on the connector is aligned
with the index point on the ring nut of
the connector port. Gently push the
handpiece connector until it is completely
inserted;
CAUTION:
To avoid damaging
the cord-handpiece, connect and/or
disconnect it, holding it by the connector
only. Never pull the cord.
1
CH 1
CH 2