64
MICROPIEZO
11.2.2
Accessible Parts of the Casing
MICROPIEZO is designed to operate in the electromagnetic environment specified below.
The client or user of MICROPIEZO should ensure that it is used in such an environment.
Phenomenon
Essential EMC
standard or test
method
Immunity test values
Electromagnetic
Environment Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV on contact
±2 kV, ±4 kV, ±8 kV,
±15 kV in air
The floor must be made of
wood, concrete or ceramic
tiles. If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF
communication devices
should not be used near any
part of the product, including
cables, except when they
respect the recommended
and calculated distances from
the equation applicable at the
frequency of the transmitter.
Magnetic field
at the mains
frequency
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Power supply frequency
magnetic fields should
have levels characteristic
of a typical location in a
commercial or hospital
environment.
a) If used, the interface between the PATIENT’S
physiological signal simulation and MICROPIEZO,
must be positioned within a 0.1 m radius of the
vertical plane of the uniform field area in the same
direction as MICROPIEZO.
b) MICROPIEZO, which intentionally receives RF
electromagnetic energy for its operation, must
be tested at the receiving frequency. The test can
be performed with other modulation frequencies
identified by the RISK MANAGEMENT PROCESS. This
test evaluates the BASIC SAFETY and ESSENTIAL
PERFORMANCE of an intentional receiver when
an environmental signal is in the passband. It is
understood that the receiver may not receive normally
during the test.
c) The test can be performed at other modulation
frequencies identified by the RISK MANAGEMENT
PROCESS.
d) Only applicable to equipment and systems with
magnetically sensitive components or circuits.
e) During the tests, MICROPIEZO can be supplied with
any NOMINAL input voltage, but with the same
frequency as the test signal.
f) Before applying the modulation.
g) This test value assumes a minimum distance between
MICROPIEZO and the magnetic field sources with
a supply frequency of at least 15 cm. If the RISK
ANALYSIS indicates that MICROPIEZO will be used
less than 15 cm from the sources of the magnetic field
with power frequency, the immunity test value should
be adjusted according to the minimum expected
distance.
Содержание MICROPIEZO
Страница 1: ...MICROPIEZO USE AND MAINTENANCE MANUAL MANUALE D USO E MANUTENZIONE MODE D EMPLOI ET D ENTRETIEN 0051...
Страница 2: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figura 1 Componenti Components Composants...
Страница 3: ...MICROPIEZO MANUALE D USO E MANUTENZIONE...
Страница 4: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figura 1 Componenti MICROPIEZO...
Страница 8: ...IV MICROPIEZO PAGINA LASCIATA INTENZIONALMENTE BIANCA...
Страница 87: ......
Страница 88: ...02150403 MICROPIEZO V IT EN FR Rev 07 del 31 05 2021...
Страница 89: ...MICROPIEZO USE AND MAINTENANCE MANUAL...
Страница 90: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figure 1 Components MICROPIEZO...
Страница 94: ...IV MICROPIEZO PAGE INTENTIONALLY LEFT BLANK...
Страница 173: ......
Страница 174: ...02150403 MICROPIEZO V IT EN FR Rev 07 del 31 05 2021...
Страница 175: ...MICROPIEZO MODE D EMPLOI ET D ENTRETIEN...
Страница 176: ...A D D1 D2 F C E B I H G J D3 D4 D5 Figure 1 Composants MICROPIEZO...
Страница 180: ...IV MICROPIEZO PAGE INTENTIONNELLEMENT LAISS E EN BLANC...
Страница 259: ......
Страница 260: ...02150403 MICROPIEZO V IT EN FR R v 07 du 31 05 2021...
Страница 261: ......