FAULT MESSAGES
8
FAULT MESSAGES
List of fault messages - self-test faults
If a self-test fault occurs during operation, it will be displayed as a plain text fault message
in the 2nd line of the LCD display of the transducer. Depending on the preset language, the
message is output in German (default) or English.
Notification on the display
Description
Possible cause of fault and
its rectification
Tube empty
Empty pipe detection is switched
on. Pipe is empty.
Ensure it is filled.
Coil current
Line interruption to the field coil. The
signal outputs are set to zero.
Send mag-flux T4 back to MECON
GmbH
Overdriven measuring circuit
The measured electrode voltage is
too large. The signal outputs are set
to zero.
Flow is too great. V >10 m/s
Power overdriven
The output of the current interface
is overdriven.
Flow is too great. V >10 m/s
ext. EEPROM is missing
The data storage module (DSB) with
the calibration data and settings of
the transducer is not plugged in.
Send mag-flux T4 back to MECON
GmbH
Tab. 12 Fault messages