10
Mecmesin
Vortex-dV Operating Manual
Jog the drive spindle clockwise at speed as set (or turn wheel clockwise)
Jog the drive spindle anticlockwise at speed as set (or turn wheel
anticlockwise)
Go back to test readiness (
A
)
E: Settings
Confirm selection (or press dial button)
Navigate up a menu selection or value (or turn wheel clockwise)
Navigate down a menu selection or value (or turn wheel anticlockwise)
Go back to previous menu item, or from Settings (
E
) to test readiness (
A
)
3.7
Settings
All settings are made by moving the selection marker to the required menu item or digit,
and confirming with the tick button or using the central dial button.
3.8
Understanding angular position
An angular position of xº means ‘rotation by xº from tared zero, in the direction of travel
according to the sign’. A set position of +270º means ‘rotated 270º from tared zero in a
clockwise direction’. An angular position greater than 360º will always produce
movement of greater than one complete rotation in relation to tared zero. The drive
spindle will therefore never take the quickest route to a position, even to a start test
position.
Содержание Vortex-dV 1.5 N.m
Страница 1: ...Vortex dV Motorised Torque Testing System Operating Manual 431 468 02 May 2018 ...
Страница 4: ...iv Mecmesin Vortex dV Operating Manual ...
Страница 21: ...Vortex dV Operating Manual Mecmesin 17 Declaration of Conformity ...
Страница 22: ...18 Mecmesin Vortex dV Operating Manual ...
Страница 23: ...Vortex dV Operating Manual Mecmesin 19 ...