CrimpTest-1 kN Calibration Checking Rig
Mecmesin
1
Overview
The CrimpTest-1 kN calibration checking rig is designed for
verifying the accuracy of the loadcell readings. For the user,
this ensures that a unit does not continue to be used with
inaccurate readings across its range. Authorised personnel,
such as Mecmesin or your distributor, can adjust the loadcell
readings, but this does not replace the need for periodic
certified recalibration.
Locally calibrated mass hangers and masses are not
supplied. Although the unit is laboratory calibrated to 1 kN,
the maximum load that can be manually applied using this
document’s method, depends on the user organisation’s
own health and safety regulations. Our maximum
recommended load is 500 N. This is referred to as the
‘maximum load’.