FUNCTIONAL DESCRIPTION: 

Pressure Release Valve (3) 

This can be used to reduce or release the pressure in the system.  Minimal force is 
required to seal the system 

Fine Adjustment Vave (4) 

The Pressure generated can be finely adjusted by turning the fine adjustment valve 
(4) either clockwise or anticlockwise to increase or decrease pressure accordingly. 

Pressure / vacuum Selector (5)  

Press the selector (5) as indicated on the label to engage the desired mode. Ensure 
that the release valve (3) is closed firmly (clockwise motion) prior to pumping. 

Knurling Thumb Nut (2) (Over pressure protection) 

To adjust the maximum output pressure of the system turns the knurling thumb nuts 
(2) to increase or decrease the stroke length so that the pressure is controlled. 

 

4.0 OPERATING PROCEDURE

 

  Connect the master / test instrument using the appropriate adapter and seals to 

the  pressure  port  at  the  end  of  the  flexible  hose  (6)  or  directly  to  the  body  (10) 

using appropriate adaptor 

  Open  the  fine  adjustment  valve  (4)  in  the  anticlockwise  direction  until  red  mark 

appears 

  Ensure the pressure release valve (3) needle tightened in the clockwise direction. 

  Pressure / vacuum calibration can be selected by Pressure / vacuum selector (5). 

  Generate pressure / vacuum by squeezing handles (1) together. Ensure that the 

handles  are  fully  squeezed  together  on  each  stroke  to  achieve  maximum 

pressure / vacuum output. 

Содержание PHP-10

Страница 1: ...AND PUMP MECHATRON CALIBRATION AND INSTRUMENT 19 37 2ND Main Road Sabari Nagar Mughalivakkam Chennai 600116 Tamilnadu India Phone 9940060420 9962060420 Mail info mechatroncalibration com Website www m...

Страница 2: ...PNEUMATIC HAND PUMP PHP 10 20 25 35 40 60...

Страница 3: ...factured with high quality for calibrating analog gauges by comparison measurements It can be used for lab and field applications There is a pressure vacuum changeover switch in the pump which is used...

Страница 4: ...d intervention and or use FEATURES Portable Light in weight Low Cost Rugged Construction SAFETY INSTRUCTIONS 1 Check for system fittings and connection tight and leak free Tighten the pressure release...

Страница 5: ...expansion of flexible hose 3 Red line indicator The Fine adjustment should not be forced to move beyond the red line indicator on the body 2 2 TECHNICAL SPECIFICATIONS Pressure Medium Air Working Pres...

Страница 6: ...PUMP ASSEMBLY...

Страница 7: ...F Swivel adaptor 7 Nylon washer 8 Plastic plug 9 Flexible hose to item under test 10 Pressure port 3 8 BSP Female connection 11 Nylon washer 3 8 ACCESSORIES STANDARD 1 Hose 1 meter with swivel adaptor...

Страница 8: ...To adjust the maximum output pressure of the system turns the knurling thumb nuts 2 to increase or decrease the stroke length so that the pressure is controlled 4 0 OPERATING PROCEDURE Connect the mas...

Страница 9: ...adjustment valve 4 for a precise pressure vacuum adjustment After the calibration completely release the pressure vacuum by twist off the vent valve Release valve gradually in anticlockwise direction...

Отзывы: