Pipe reducers
If the pipe diameter is not the same as the diameter of sensor, then pipe reducers can be used. So as not to
lose accuracy of the measurement, the slope of reducers should not exceed 8
.
Vertical mounting
When the sensor is mounted on a vertical section of pipe, the flow direction must be upwards. In the case of
a downward flow direction, air bubbles could collect in the sensor resulting in unstable and inaccurate
measurement.
Pumps
Never install the sensor on the suction side of a pump or on a section of pipe where a vacuum is possible.
8