Chapter 3
52
Connecting Digital Output Signals
shows how to connect a digital output (line 0, in this case) to the STP25 screw
terminal panel. Refer to
for the screw terminal assignments of the STP25
screw terminal panel.
Figure 22: Connecting Digital Outputs
Digital Out 0
Digital Ground
16
17
STP25
500
Ω
1 Out = LED On
From
DT9857
or
DT9857E
100
Ω
The output current is determined using the following equation:
In this example, if the maximum output voltage is 3.3 V, the LED voltage drop
is 0.7 V, the internal resistor is 100
Ω
, and the external resistor is 500
Ω
, the
maximum output current is 4.33 mA. Using the minimum output voltage of
2.4 V with the same resistor values, the minimum output current is 2.83 mA.
Current
Out
Voltage
Out
Voltage
–
LED
R
Internal
R
External
+
---------------------------------------------------------------------
=
Содержание Data Translation DT9857
Страница 1: ...DT9857 and DT9857E UM 25656 M User s Manual Title Page ...
Страница 4: ......
Страница 10: ...Contents 10 ...
Страница 14: ...About this Manual 14 ...
Страница 15: ...15 1 Overview Hardware Features 16 Supported Software 19 Supported Accessories 22 Getting Started Procedure 24 ...
Страница 25: ...Part 1 Getting Started ...
Страница 26: ......
Страница 40: ...Chapter 2 40 ...
Страница 61: ...61 4 Verifying the Operation of a Module Select the Device 63 Acquire Data from an Analog Output Channel 65 ...
Страница 64: ...Chapter 4 64 ...
Страница 71: ...Part 2 Using Your Module ...
Страница 72: ......
Страница 126: ...Chapter 6 126 ...
Страница 127: ...127 7 Troubleshooting General Checklist 128 Technical Support 130 If Your Module Needs Factory Service 131 ...
Страница 132: ...Chapter 7 132 ...
Страница 140: ...Chapter 8 140 ...
Страница 180: ...Appendix B 180 ...
Страница 181: ...181 C Register Level Programming Writing to the EEPROM Register 182 ...
Страница 184: ...Appendix C 184 ...
Страница 192: ...Index ...