Chapter 9
176
General Checklist
Should you experience problems using the DT3010 Series board, follow these steps:
1.
Read all the documentation provided for your product. Make sure that you have added
any “Read This First” information to your manual and that you have used this
information.
2.
Check the OMNI CD for any README files and ensure that you have used the latest
installation and configuration information available.
3.
Check that your system meets the requirements stated in
the README file on the OMNI CD.
4.
Check that you have installed your hardware properly using the instructions in
5.
Check that you have installed and configured the device driver properly using the
.
6.
Search the DT Knowledgebase in the Support section of the Data Translation web site (at
www.
mccdaq
.com) for an answer to your problem.
If you still experience problems, try using the information in
to isolate and solve the
problem. If you cannot identify the problem, refer to
.
Table 28: Troubleshooting Problems
Symptom
Possible Cause
Possible Solution
Board does not
respond.
The board configuration is
incorrect.
Check the configuration of your device driver to
ensure that the board name and type are correct
.
The board is incorrectly aligned in
a PCI expansion slot.
Check that the slot in which your DT3010 Series
board is located is a PCI slot and that the board is
correctly seated in the slot
.
The board is damaged.
Contact Data Translation for technical support; refer
to
Intermittent
operation.
Loose connections or vibrations
exist.
Check your wiring and tighten any loose connections
or cushion vibration sources.
The board is overheating.
Check environmental and ambient temperature;
consult the board’s specifications on
of this
manual and the documentation provided by your
computer manufacturer for more information.
Electrical noise exists.
Check your wiring and either provide better shielding
or reroute unshielded wiring.
Data appears to
be invalid.
An open connection exists.
Check your wiring and fix any open connections.
A transducer is not connected to
the channel being read.
Check the transducer connections.
The board is set up for differential
inputs while the transducers are
wired as single-ended inputs or
vice versa.
Check your wiring and ensure that what you specify
in software matches your hardware configuration
.
Содержание Data Translation DT3010 Series
Страница 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Страница 5: ......
Страница 15: ...About this Manual 14 ...
Страница 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Страница 23: ...Chapter 1 22 ...
Страница 24: ...Part 1 Getting Started ...
Страница 25: ......
Страница 37: ...Chapter 2 36 ...
Страница 49: ...Chapter 3 48 ...
Страница 91: ...Chapter 4 90 ...
Страница 102: ...Part 2 Using Your Board ...
Страница 103: ......
Страница 147: ...Chapter 6 146 ...
Страница 159: ...Chapter 7 158 ...