Figure 78: Location of Potentiometers R12 to R15 on the DT3010, DT3010-268, DT3010/32,
and DT3010/32-268
9.
In the Voltages box,
9.3750
.
10.
Physically adjust potentiometer R14 (labelled 0G) on the DT3010, DT3010-268,
DT3010/32, or DT3010/32-268 board until the display reads +9.3750 V (within 0.001 V).
shows the location of this potentiometer.
11.
In the D/A box, select
DAC 1
.
12.
In the Voltages box, select
–9.3750
.
13.
Physically adjust potentiometer R13 (labelled 1Z) on the DT3010, DT3010-268,
DT3010/32, or DT3010/32-268 board until the display reads
–
9.3750 V (within 0.001 V).
shows the location of this potentiometer.
14.
In the Voltages box, select
+
9.3750
.
15.
Physically adjust potentiometer R12 (labelled 1G) on the DT3010, DT3010-268,
DT3010/32, or DT3010/32-268 board until the display reads +9.3750 V (within 0.001 V).
shows the location of this potentiometer.
Note:
If you want to check the values for intermediate ranges, select
Display Values
in
the Mode box and select any of the available ranges; the range is then displayed.
You
cannot calibrate intermediate ranges
.
16.
Click
Quit
when you are finished calibrating the analog output circuitry.
Once you have finished this procedure, the analog output circuitry is calibrated. To close the
DT3010 Calibration Utility, click the close box in the upper, right corner of the window.
R12 -1G
R13 -
1Z
R14 - 0G
R15 -
0Z
Potentiometers for the DACs
Содержание Data Translation DT3010 Series
Страница 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Страница 5: ......
Страница 15: ...About this Manual 14 ...
Страница 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Страница 23: ...Chapter 1 22 ...
Страница 24: ...Part 1 Getting Started ...
Страница 25: ......
Страница 37: ...Chapter 2 36 ...
Страница 49: ...Chapter 3 48 ...
Страница 91: ...Chapter 4 90 ...
Страница 102: ...Part 2 Using Your Board ...
Страница 103: ......
Страница 147: ...Chapter 6 146 ...
Страница 159: ...Chapter 7 158 ...
Страница 160: ...159 8 Calibration Calibrating the Analog Input Subsystem 161 Calibrating the Analog Output Subsystem 169 ...
Страница 176: ...175 9 Troubleshooting General Checklist 176 Technical Support 178 If Your Board Needs Factory Service 179 ...
Страница 181: ...Chapter 9 180 ...
Страница 193: ...Appendix A 192 ...
Страница 213: ...Appendix B 212 ...
Страница 231: ...Index 230 ...