case of temporary interruption of the main power supply. Mdoloris Medical Systems
recommends using the ANI Monitor V1 without power supply (battery mode) for 15 minutes
maximum.
Check the battery charge symbol before unplugging the main power supply; if the internal
battery is insufficiently charged, the ANI Monitor V1 will switch off automatically.
After reconnecting the ANI Monitor V1 to the main power supply, press the power switch on
the front side of the monitor. In case the monitor does not start, press the switch at the back of
the monitor in order to reboot the device.
Never use the ANI Monitor V1 only on battery power during surgery. The monitor must be
plugged on the main power supply, especially while an electric knife is being used.
Only personnel trained by Mdoloris Medical Systems can safely perform service or repairs on
ANI Monitor V1. However, the following elements may be replaced by personnel untrained in
technical maintenance (following the manufacturer's instructions):
-
End User Cable;
-
Power cord;
-
Acquisition unit (BA-ANI-V1);
-
Pole clamp.
This medical equipment, its components and packaging must be recycled in accordance with
local regulations on the environment and disposal of electric waste.