4.3 TIME MEASUREMENTS

4.3.1 TIME SPACE MEASUREMENTS

This is a procedure for making time (period) measurements between two points on a
waveform:
(1)

 

Connect the signal to be measured to the input terminal [18].

(2)

 

Adjust level to obtain steady waveform.

(3)

 

Turn VAR clockwise to the calibration position, and set sweep controls to obtain a
normal display of 1-2 signal cycles.

(4)

 

Using the vertical and horizontal positions, set two points to be measured in the
waveform on the same horizontal level.

(5)

 

Measure the distance between the two points , the measurement is calculated by the
following equation:

 Distance between 2 points (DIV) 

×

 rate of sweep (t/Div)

TIME (S) = --------------------------------------------------------------------

Horizontal factor

Shown in figure 4-6, distance between A & B is 8 Div. sensitivity is 2 

µ

S/Div, Horizontal

factor x 1, then Time measurement is 16

µ

S

Fig 4-6

4.3.2

 

CYCLE & FREQUENCY MEASUREMENTS

Shown in Fig 4-6, frequency measurements are made by measuring the time period of

one cycle of waveform (T), and calculating the frequency that equals the reciprocal of
the time period. For example, T=16

µ

S, then frequency is:

   

    1

F =1/T = ------------------ = 62.5 KHz

  16 

×

10

-6

4.3.3 PULSE RISE TIME AND FALL TIME MEASUREMENTS

A

B

Содержание CQ5010B

Страница 1: ...CQ5010B SINGLE TRACE 10MHZ OSCILLOSCOPE USER S MANUAL CONTENTS 1 SAFETY PRECAUTION 2 SPECIFICATIONS 2 1 VERTICAL SYSTEM 2 2 TRIGGER SYSTEM...

Страница 2: ...nsitivity up to 50V div Sweep rate at 0 1S DIV 0 1 S DIV on horizontal system The oscilloscope is easy to operate and highly reliable It is an ideal instrument for research production education and de...

Страница 3: ...otten into this instrument first switch the power off remove the power cord and call for repair z Do not place this instrument on an unstable place such as on a shaky stand or on a slant The dropping...

Страница 4: ...Range 05 V 2 Frequency 1kHz 2 2 6 CRT Display Area 8 10DIV 1DIV 6mm Accelerating Voltage 1200V Display Color Green 2 7 POWER SOURCE Voltage Range 110V 10 Frequency 60Hz 2Hz Power Consumption 25W 2 8...

Страница 5: ...2 1 50 20 10 5 V mV VOLTS DIV POSITION POSITION VOLT VAR CAL CAL TIME VAR CAL CAL LEVEL _ PROBE ADJUST FOCUS INTENSITY POWER 10MHz HANDY OSCILLOSCOPE AUTO NORM TV INT LINE EXT Y INPUT 1M 30pF 1M _ X E...

Страница 6: ...gger level in a negative going direction EXT X X Y switch 11 AUTO NORM TV AUTO a single trace shown on screen even no signal Automatically reverts to triggered sweep operation when adequate triggered...

Страница 7: ...stment a Slightly turn 20 22 counter clockwise but not to unscrew b Insert a straight end screw driver to 21 and observe the wave move screw driver in order to let the waveform parallel with the horiz...

Страница 8: ...level as starting reference 3 3 6 SIGNAL CONNECT 1 Probe operating Use 10 1 to 1 1 alternating probes When using 10 1 probe input impedance is 10M 16 pF While if 1 1 is used for observing lower signa...

Страница 9: ...TS DIV and observe waveform let waveform display on the screen within 5 divisions and turn VAR clockwise to the calibration position 3 Adjust level to make waveform steady 4 Adjust sweep controls to s...

Страница 10: ...LTS DIV let waveform show in the proper position on the screen turn VAR to the calibration position 6 Read the divisions between the zero reference level to the wave form by the tested object 7 Calcul...

Страница 11: ...eform on the same horizontal level 5 Measure the distance between the two points the measurement is calculated by the following equation Distance between 2 points DIV rate of sweep t Div TIME S Horizo...

Страница 12: ...o coincide with a vertical reference line 6 Measure the horizontal distance in divisions between the 10 and 90 points on the waveform divisions 7 Pulse rise time and fall time measurement is calculate...

Страница 13: ...of sweep signals signals of Lissajous pattern or other equipment s display setup X Y mode operation is turn 10 to EXT X input signals through 13 sensitivity to be adjusted directly with 14 then input...

Отзывы: