![McIlwain Tissue Chopper Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/mcilwain/tissue-chopper/tissue-chopper_instructions_1754949002.webp)
Instructions for
The McIlwain Tissue Chopper
IMPORTANT
The McIlwain Tissue Chopper must be fitted with a three-pin plug and only connected to an A.C. supply
having a voltage corresponding to that marked on the appliance. When a fused plug is used it should be fitted
with a 3 amp fuse, if in doubt consult a qualified electrician, since these instructions must be observed.
The wires in the mains lead supplied with this appliance are coloured in accordance with the following code.
Green and Yellow
:
Earth
Blue
:
Neutral
Brown
: Live
As these colours may not correspond with coloured markings identifying the terminals in your plug, connect as
follows:
The Green and Yellow to the terminal marked with the letter E or by the earth symbol or coloured green and
yellow.
The Blue wire to the terminal marked N or coloured blue.
The Brown wire to the terminal marked L or coloured brown.
WARNING: THIS APPLIANCE MUST BE EARTHED
Use the Tissue Chopper on a firm bench and do not restrict ventilation by using in a confined space.
Clean the Tissue Chopper with a dry cloth, we do not recommend the use of solvents. To sterilise the blade
holder, cutting table and plastic disc see “Operating Procedure”.
If an object is accidentally dropped into the Tissue Chopper, immediately disconnect from the main electricity
supply and retrieve the object. If the equipment has been dropped or damaged in any way, it should be checked
before connecting to the mains supply. Similarly, although the equipment is lubricated for life, any work
[especially involving the removal of covers] must be carried out by a competent Service Engineer.
In the interest of safety, always use the equipment in the specified manner.
BY NATURE OF ITS USE, THIS INSTRUMENT DOES NOT HAVE A GUARD AROUND THE
CUTTING BLADE. IN THE INTEREST OF SAFETY, CARE MUST THEREFORE BE
EXERCICESD WHEN USING THIS EQUIPMENT. AT ALL TIMES KEEP HANDS CLEAR OF THE
BLADE AND CUTTING TABLE WHEN THE INSTRUMENT IS SWITCHED ON. WE
RECOMMEND THE BLADE IS REMOVED WHEN THE APPARATUS IS NOT IN USE.