McGan MM513 Скачать руководство пользователя страница 2

10.

 

Follow the following steps: 

 

a.

 

Push the ESI under test through the LSE ring electrode slowly (approximately 3 feet every 4 seconds)  

b.

 

The Alarm will sound when the ESI is first inserted into the electrode as that is the bare tip of the instrument. 

c.

 

Alarm will sound and LED will Flash if a fault is found in the coating which will indicate a fault with the instrument. 

 

 

11.

 

After the test is completed turn the base unit off and remove the clamp end from the unit under test, remove the electrode from the probe wire and 
remove the ground wire and probe wire from the base unit. Properly store the unit and accessories away.   

 

12.

 

Follow the proper established hospital procedure after testing is completed with regards to the instrument under test. 

 

 

NOTES:  

A.

 

The unit should always be switched off prior to removing or repositioning of the ground lead, the HV red wire or the Saddle Block.. 

B.

 

If the unit is on and you touch the ground lead (clamp end) and the probe end of the base unit at the same time you will rece ive a very mild 

“tingle”. To remove the possibility of receiving the “tingle” always use surgical gloves when handling the leads. 

C.

 

You can hold the Saddle Block from the top or the sides as long as you do not touch the connection points. 

 

 

 

 

 

 

 

 

 

 
 
 

Note: Using the Tri-Hole Electrode: 

 

 

The Set-up is the same as shown above except turn the voltage to 4.2KV +/- .3KV 

 

a.

 

Insert the round electrosurgical instrument into hole size closest to the diameter of the ESI under test. NOTE: Hole sizes are slightly 
larger than 3mm, 5mm and 10mm from the bottom (pin side) up

.

 

 

For Bi-Polar Instruments: (wear gloves) 
 

a.

 

The Set-up is the same as the steps in #5 

b.

 

Place the Brush Electrode into the Saddle Block in the slot on the right side away from the pin. 

 
 
 
 
 
 
 
 
 
 
 

a.

 

Attach the Green ground wire to the back end of the bi-Polar forceps. Make sure the clamp is connected to both pins. 

b.

 

Insert the end of one tine of the Bi-Polar forceps into the middle of the brush, 

 
 
 
 
 
 
 
 
 
 
 

Hold on the top  

Hold on the 
sides of the 
base 

Содержание MM513

Страница 1: ...nit until the suction feet stick to the surface 4 There are a number of ways to set the Saddle Block up depending upon the electrosurgical instrument ESI to be tested and or the McGan kit used 5 Attach the red port on the top of the MM513 unit directly to the side pin of the Saddle Block Make sure controls face up 6 Insert the chosen electrode securely into the proper slot on the Saddle Block 7 Ta...

Страница 2: ...If the unit is on and you touch the ground lead clamp end and the probe end of the base unit at the same time you will receive a very mild tingle To remove the possibility of receiving the tingle always use surgical gloves when handling the leads C You can hold the Saddle Block from the top or the sides as long as you do not touch the connection points Note Using the Tri Hole Electrode The Set up ...

Страница 3: ... unit red port optional 1 Attach the clamp end of the Green ground wire to conductive core of instrument under examination See Picture A B and C for illustrations Using the Optional HV RED Lead PN MMRWP 006 not included in kit 1 Insert the Red HV lead probe wire firmly into the red port top of the base unit 2 Insert wire saddle brush ring or Tri hole electrode firmly into end of yellow handle and ...

Страница 4: ...0 Base unit MM513 120 AC battery charger MMGWC 0005L Ground wire green with Large alligator clip MMRWP 0006 HV lead red with mini handle MMBRU 0007 8mm Brush electrode MMLSE 0029 LS Ring Electrode MMTRI 0022A Tri Hole Electrode MMSBT 170 Saddle Block Only MM513 100 Kit contains the base unit battery AC charger HV and ground leads brush ring Tri Hole electrodes Saddle Block Quick Start Manuals Trai...

Отзывы: