McGan MM513 Скачать руководство пользователя страница 1

Attach ground 
clamp here 

The LSE and Tri-Hole 
Electrodes mounted in 
Saddle Block 

 

MM513 

QUICK OPERATION GUIDE 

 

CAUTION: Please read the full MM513 manual in detail before operating the unit. Review all warnings 
noted in the manual. 

Use:

 MM513 is a low frequency high voltage insulation defect tester seeking crack and pinholes 

in the jacket or coating of laparoscopic and bi-polar electrosurgical instruments. 

Description: The MM513 system is a non-destructive, non-patient contact, high voltage insulation tester 
designed to test the insulation integrity of electrosurgical instruments.   
It should only be used 

i

the Central Sterile a

r

e

a ONLY.

 

 

 

1.

 

Remove the MM513 unit and accessories from the carrying case. 

2.

 

Take the Green ground wire and firmly insert it into the green port on the bottom of the base unit. 

 

 

 
 
 
 
 

 
 
 

3.

 

Secure the Saddle block to a flat, preferably metal surface, by pushing the top of the unit until the suction feet stick to the surface.  

4.

 

There are a number of ways to set the Saddle Block up depending upon the electrosurgical instrument (ESI) to be tested and/or the McGan kit used. 

5.

 

Attach the red port on the top of the MM513 unit directly to the side pin of the Saddle Block.. Make sure controls face up 

 
 
 
 
 
 
 
 
 
 
 
 

6.

 

Insert the chosen electrode securely into the proper slot on the Saddle Block 

7.

 

Take the clamp on the green ground wire and attach it to the conductive core of the instrument under test.  
 
 
 
 
 
 
 
 
 
 

8.

 

Turn the base unit on and set the voltage to 2.8KV +/- 0.3KV  

9.

 

Check Battery LED indication colors: 

Red = Battery Flat     Blue = Charging   Green = Battery Full

 

a.

 

If battery level is blue or red recharge unit using the adapter supplied with the MM513  kit  

b.

 

use of any other charger may cause damage to the MM513 unit and void warranty

 

 

CAUTION:

  

DO NOT

 simultaneously handle the brush electrode and ground clamp as it will cause a mild “tingle”. 

Use surgical  

gloves as a precaution against receiving the “tingle”.

  

 
 
 

 

MM513 
attached directly 
to pin  

Saddle Block 
 
 
Tri Hole electrode 
Brush Electrode 

LSE Electrode 

Содержание MM513

Страница 1: ...nit until the suction feet stick to the surface 4 There are a number of ways to set the Saddle Block up depending upon the electrosurgical instrument ESI to be tested and or the McGan kit used 5 Attach the red port on the top of the MM513 unit directly to the side pin of the Saddle Block Make sure controls face up 6 Insert the chosen electrode securely into the proper slot on the Saddle Block 7 Ta...

Страница 2: ...If the unit is on and you touch the ground lead clamp end and the probe end of the base unit at the same time you will receive a very mild tingle To remove the possibility of receiving the tingle always use surgical gloves when handling the leads C You can hold the Saddle Block from the top or the sides as long as you do not touch the connection points Note Using the Tri Hole Electrode The Set up ...

Страница 3: ... unit red port optional 1 Attach the clamp end of the Green ground wire to conductive core of instrument under examination See Picture A B and C for illustrations Using the Optional HV RED Lead PN MMRWP 006 not included in kit 1 Insert the Red HV lead probe wire firmly into the red port top of the base unit 2 Insert wire saddle brush ring or Tri hole electrode firmly into end of yellow handle and ...

Страница 4: ...0 Base unit MM513 120 AC battery charger MMGWC 0005L Ground wire green with Large alligator clip MMRWP 0006 HV lead red with mini handle MMBRU 0007 8mm Brush electrode MMLSE 0029 LS Ring Electrode MMTRI 0022A Tri Hole Electrode MMSBT 170 Saddle Block Only MM513 100 Kit contains the base unit battery AC charger HV and ground leads brush ring Tri Hole electrodes Saddle Block Quick Start Manuals Trai...

Отзывы: