D
URA
M
AG
1.
Flow Meter Installation
Wafer Style Meter Installation
Install the Dura Mag between two flanged end pipes.
Hardware and gaskets are provided, but customers must
supply the flanged pipe ends.
Remote Mount Installation
You will need to prepare the location where you will install the remote converter. The location cannot be further from the
flow meter than the length of the 25’ cable. This must be planned in advance because the cable cannot be lengthened.
Doing so will alter the calibration accuracy between the meter and the converter and void the warranty.
Install the flow meter as required. Mount the converter and connect the cable to the meter’s junction box and the
converter’s external connection.
Flanged Meter Installation
Install the Dura Mag flow meter inline between two flanged end pipes. The flow meter may require grounding, depending
on the environment they are being installed in. Refer to section 4 for a full description of grounding methods that are
available.
Shown are customer’s
pipe ends and flanges
See section 3 for
connection
See section 3 for
connection
AC or DC
power source
See section 3 for
connection
I
IMPORTANT: The example shown above does not include grounding
installation. Depending on your requirements, the meter will need to be
grounded according to the methods shown in section 4.
The Dura Mag requires grounding to ensure proper functioning and accuracy. See section 4 for grounding
instructions and methods.
I
30122-61 Rev. 1.4 | 04OCT2021
Page 2