MBA200  

·  

Operating Instructions  8011417  V 2.2  

·  

© MBA Instruments GmbH Germany 

 

 

 

 

10 

 

 

7.2

 

Electrical data

 

Mains voltage:

 

 

 

see nameplate

 

 

Tolerance:

 

see text in this manual

 

 

AC mains frequency:

 

50 

… 60 Hz

 

 

Power  consumption

 

 

 

– without  heating: 
– with heating:

 

3 VA 
10 VA

 

 

Response  time: 
– quick-rotating  version:

 

3 s 
0,6 s

 

 

Bin-level Indicator MBA200 

 

 

6  Fault indication

 

Operation principle of the fault status indication

 

MaihakMBA   type   220   and   230   are   equipped   with   the 
»Status«  switch  contact  for fault  indication,  which  is driven  by 
relay  K2.  During  normal  operation,  K2 is activated;  in case  of 
an  internal  fault,  K2  is  deactivated.  This  also  happens  during 
power  supply  failure.  Please  note:  In  fault  condition,   also 
relay  K1  (which  drives  the  »Signal«  contact)  is  deactivated, 
in  order  to  provide  a fail-safe  operation;  however,  this  is  only 
achieved   if   the   K1   function   mode   is   properly   selected 
(

page 9, § 3.5

Fault indication caused by action monitoring

 

If    action    monitoring    is    built-in    (option),    the    following 
conditions  will trigger the fault status indication: 
–    The shaft does not rotate, although  the motor is not in the 

stop position  (switch  S2). 

Possibly  defective:  switch  S2, motor,  rotation  sensor. 

–    The  shaft  is  rotating,  although  the  motor  is  in  the  stop 

position  (switch  S2). 

– Possibly  defective:  switch S2. 

 

Remedy:  Check  if internal  micro-switches  are blocked. 

 

If this does  not help: 

Replace  the unit head. 

 

General causes for fault indication

 

Without the action monitoring  option, the following  conditions 
will trigger the fault status indication: 

●    

Power  supply  has failed. 

●    

An internal  fuse is defective. 

●    

Action  monitoring  is  activated  although  it  is  not  built-in. 

Remedy:  Correct  the setting of S1-4 (

page 9, § 3.5

). 

7  Technical data

 

 

7.1  

Specifications on the nameplate

 

●    

Unit type (for example,  »210«)  and version  code 

●    

Required  mains power supply (for example,  »230VAC  50/ 
60 Hz«) 

●    

Explosion   protection   class   (for   example,   »II   1D  

…«), 

including   the  enclosure   protection   class  (for  example, 
»IP 65«) and maximum  surface  temperatures: 
–    The first temperature  value applies to the components 

protruding  into the container. 

–    The  second   temperature   value   applies   to  the  unit 

head. 

●    

Allowable  ambient  temperature  (T

a

)  and  »Ex«  zone  (for 

example,  »Z20«): 
»extern«  = for the unit head 
»medium«  = for components  protruding into the container 

 

Specifications  on the nameplate  apply with higher priority. 

 

 

Important  specifications  on the nameplate  (example) 

 

MBA Instruments

 

MBA 210YAA1N1A00300AX

 

II 1D Ex tD A20 IP65 T100°C/T100°C

 

 

7.3  

Ambient conditions

 

Ambient  temperature  outside the container 
– without heating:  

–15 … +60 °C (5 … 140 °F) 

– with heating:  

–45 … +60 °C (–49 … +140 °F) / 

PN 214000-G 123456

 

SN 654321 05.03

 

TÜV 08 ATEX 554820 X 

T

a  

Z20/extern: 

–15°C... +60°C

 

–30 … +60 °C (–22 … +140 °F) 

1

 

230VAC 50/60Hz 10VA 

T

a  

Z20/medium:      –30°C... +80°C 

QA 

0044 

MBA Instruments 

D-25451 Quickborn 

Made in Germany 

 

 

 

 

Nameplate  stating the Certificate of Conformity  (example) 

 

MBA Instruments

 

MBA 210YAA1N1A00300AX

 

Allowable  temperature  inside the container 
– standard  versions:  

–30 … +80 °C (–22 … +176 °F) 

– special versions:  

see nameplate  (

§ 7.1

Max. surface  temperature 
of the MBA unit:  

see nameplate  (

§ 7.1

 

7.4  

Product materials

 

Product materials in contact with bulk material

 

Enclosure,  protective  tube: steel/painted  steel/stainless 

steel

2

,  aluminium

3

 

Sealing:  

Viton or PTFE

or graphite

PN 214000-G 123456

 

SN 654321 05.03

 

II 1D Ex tD A20 IP65 T100°C/T100°C 

TÜV 08 ATEX 554820 X 

T

a  

Z20/extern: 

–15°C... +60°C

 

Materials of the unit head

 

Enclosure  body:  

aluminium 

230VAC 50/60Hz 10VA 

T

a  

Z20/medium:      –30°C... +80°C

 

QA 

Enclosure  cover:  

painted  aluminium 

0044 

MBA Instruments 

D-25451 Quickborn 

Made in Germany 

 

Nameplate  stating the EC-type Examination  Certificate  (example) 

 

MBA Instruments

 

MBA 210YAA1N1A00300AX

 

Enclosure  sealing:  

silicone  (VMQ) 

Cable glands:  

PA (sealing:  NBR) / option: 

nickel-plated  brass (sealing: 
NBR/FPM,  PA/PVDF) 

Sealing  plug:  

PA 

Balancing  membrane:  

polyester/ePTFE 

(sealing:  silicone) 

PN 214000-G 123456

 

SN 654321 05.03

 

II 1D Ex tD A20 IP65 T100°C/T100°C 

TÜV 08 ATEX 554821 X 

T

a  

Z20/extern: 

–15°C... +60°C

 

230VAC 50/60Hz 10VA 

T

a  

Z20/medium:      –30°C... +80°C 

QA 

0044 

MBA Instruments 

D-25451 Quickborn 

Made in Germany 

in explosion-hazardous  locations  (

page 13, § 7.6

depending  on product  version 

standard  unit head only 

350 °C (660 °F) version 

 
 
 
 
 
 
 

 

Содержание MBA200

Страница 1: ...Operating instruction MBA200 Rotating Bin Level Indicator Installation Operation Maintenance ...

Страница 2: ...quipment 4 2 4 Appropriate use 4 2 5 Responsibility of the user 4 3 Installation 5 3 1 Protective roof deflector 5 3 2 Assembly if required 5 3 3 Installation 6 3 3 1 Operational conditions 6 3 3 2 Installation position 6 3 3 3 Attachment 6 3 3 4 Temperatures of the distance tube 7 3 4 Electrical connection 7 3 4 1 Important safety notes 7 3 4 2 Safe electrical installation 7 3 4 3 Protection agai...

Страница 3: ...Bin level Indicator MBA200 MBA200 Operating Instructions 8011417 V2 2 MBA Instruments GmbH Germany 2 1 Product overview ATEX Zone 20 21 Option 80 C 200 C 500 C 800 C 80 C 200 C 350 C ...

Страница 4: ...again The motor mechanism requires approximately 3 seconds to indicate stopping or restarting of the paddle switching delay due to the mechanical sequence Additional indication delay can be effected by the characteristics of the bulk material torsion effects e g cable shaft rubber paddle electronic switching delay page 9 3 5 2 3 Optional equipment General options Special seal DTR Protection agains...

Страница 5: ...y attach both parts of the protective tube Recommendation Now make two small countersunk holes in the coupling sleeve which will fix the position of the locking screws use a max 3 2 mm diameter drill guided through the screw holes of the coupling sleeve Put in the locking screws and tighten Install the protective tube for units with a protective tube Apply a threadlocking adhesive such as Loctite ...

Страница 6: ...9 Provide a suitable water dust sealing between device and container If the paddle does not go through the container opening Detach the paddle before mounting and re install after mounting Height adjustment option The height adjustment device is a clamp ring fitting which allows to fix the protective tube at variable insertion depth 1 Install the threaded base part of the fitting in the container ...

Страница 7: ... head for service purposes Take care Give a warning to other persons 3 4 2 Safe electrical installation A MBA200 should be installed and set up by skilled persons who can assess the tasks given and consider the dangers involved The MBA200 is not equipped with a power switch or fuse Install an external power switch which can switch on and off the mains power supply to the MBA200 Provide an external...

Страница 8: ...age specification on the nameplate page 11 7 1 Overheat fuse Type 210 is equipped with an overheat fuse which cuts off the power supply when the internal temperature exceeds 98 C 208 F When the fuse is blown the electronics board needs to be replaced for repair Wiring example as FULL level indicator Signal contact Use make break relay contact K1 for the level indication paddle rotating stopped Dep...

Страница 9: ...o terminal 0 Connect the protective earth PE conductor to the terminal in the unit head WARNUNG Internally generated power voltage The power terminals 230 and 115 will have a voltage output of 230 115 V even if the unit is used with a lower power voltage due to power feedback through internal transformer Observe this information when operating the unit with open unit head for service purposes Take...

Страница 10: ... which protrude into the container shaft paddle etc Pay special attention to the connecting parts split pins etc Replace damaged or dubious parts Clean the protective tube only if the unit is equipped with an open protective tube Clean the inside of the protective tube to make sure that the shaft can always freely rotate Clean the distance tube if existing if required for explosion protection page...

Страница 11: ...st temperature value applies to the components protruding into the container The second temperature value applies to the unit head Allowable ambient temperature Ta and Ex zone for example Z20 extern for the unit head medium for components protruding into the container Specifications on the nameplate apply with higher priority Important specifications on the nameplate example MBA Instruments MBA 21...

Страница 12: ...0 18 16 DN 100 PN 16 220 180 18 20 DN 125 PN 6 240 200 18 18 DN 125 PN 16 250 210 18 22 4 ANSI Class 150 228 6 190 5 19 5 ANSI Class 150 254 5 215 9 22 2 L 125 118 28 29 125 L 178 D 60 70 Bin level Indicator MBA200 7 5 Dimensions 160 58 A 41 C 48 G1 A B C D 33 2 33 2 mm L Eintauchtiefe L D 220 ...

Страница 13: ...MBA200 Operating Instructions 8011417 V2 2 MBA Instruments GmbH Germany 12 Bin levellndicator MBA2CO 7 6 EC Type Examination Certificate Page 1 of 2 ...

Страница 14: ...Bin level Indicator MBA200 Notes ...

Страница 15: ...8011417 2 2 2015 10 MBA200 MBA Instruments GmbH Friedrich List Straße 7 25451 Quickborn Germany Phone 49 4106 123 88 80 www mba instruments de info mba instruments de ...

Отзывы: