» Never install the sensor in a location that will expose the sensor to patient exhalation or secretions.
» Only use Maxtec approved accessories and replacement parts. Failure to do so may seriously  
   impair the MaxVenturi’s performance. Repair or alteration of the MaxVenturi beyond the scope  
   of the maintenance instructions or by anyone other than authorized Maxtec service personnel 
   could cause the product to fail to perform as designed.
» Use of the MaxVenturi near devices that generate strong electrical fields may cause erratic readings.
» This device has a visual low-battery alarm but no audible alarm.
» Calibrate the MaxVenturi weekly when in operation or if environmental conditions change significantly,  
   i.e., Temperature, Humidity, Barometric Pressure. (refer to Calibration sections of this manual).
» The device will assume a percent oxygen concentration when calibrating. Be sure to apply
   100% oxygen or room air concentration to the device during calibration or the device will not
   calibrate correctly. (See section 2.2)

 DO NOT

 use ethylene oxide sterilization.    

DO NOT

 immerse the device in any cleaning solution,

 autoclave or expose the device to high temperatures.

» Use with any other patient interface system may result in false readings from the flow meter.
» The MaxVenturi is not intended for steam, ethylene oxide or radiation sterilization.

 DO NOT

 clean with ethanol or acetone.

» After cleaning, and before use on a patient, attach the device to an oxygen supply and allow the
   device to run at high flow for several minutes to allow any cleaning fluids or vapors to evaporate
   and be flushed out.
» The device can flow excess oxygen out the entrainment port if the oxygen control knob is
   turned up too high. This can lead to a minor drop in total air flow to the patient and excess
   oxygen entering the room environment.
» Federal Law (USA) restricts this device to sale by or on the order of a physician.

ATTENTION:

 Indicates a potentially hazardous situation, if not avoided, could result in minor or 

moderate injury and property damage.

» If the MaxVenturi is ever exposed to liquids (spills or immersion) or to any other physical abuse,        
   return to Maxtec for evaluation before use.
» Always remove the batteries to protect the unit from potential leaky battery damage when the 
   unit is going to be stored or not in use for periods exceeding 1 month. Replace dead batteries 
   with high quality AA Alkaline batteries.

 DO NOT

 use rechargeable batteries with this device.

» Maxtec cannot warranty any damage resulting from misuse, unauthorized repair or improper  
   maintenance of the instrument.
» This product is Latex free.
» Avoid usage in environments with greater than 95% relative humidity.

Содержание MaxVenturi R211P03

Страница 1: ...O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R211M03 Rev Q R 2 1 1 P 0 3 MaxVenturi A B C D E F 60 50 40 30 20 10 02...

Страница 2: ......

Страница 3: ...ars from the date of shipment from Maxtec Based on Maxtec s product evaluation Maxtec s sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equ...

Страница 4: ...l times by the filter listed in the disposables list The inlet filter prevents entrainment of ambient contamination and silences venturi noise The filter provided with the MaxVenturi is single patient...

Страница 5: ...ult in false readings from the flow meter The MaxVenturi is not intended for steam ethylene oxide or radiation sterilization DO NOT clean with ethanol or acetone After cleaning and before use on a pat...

Страница 6: ...ent Interfaces 7 2 5 3 Inlet Filter 9 3 0 FACTORS INFLUENCING ACCURATE READINGS 9 3 1 Elevation Pressure Changes 9 3 2 Temperature Effects 9 4 0 CALIBRATION ERRORS AND ERROR CODES 9 5 0 CHANGING THE B...

Страница 7: ...2 11 10 Flow meter Oxygen On Off Valve Inlet Oxygen Fitting Oxygen Control Knob O2 Flow Control Knob Room Air Inlet Room Air Inlet Filter ON OFF Button Calibration Button MAX 250 Series Oxygen Sensor...

Страница 8: ...y the humidifier patient interface system The numbered scale on the flow meter that corresponds to the compensated flow rate for the Fisher Paykel Optiflow System and Hudson RCI Neptune Heated Humidif...

Страница 9: ...displayed O2 percentage see Factors Influencing Accurate Readings in section 3 0 The MaxVenturi can be calibrated at 100 oxygen or room oxygen 20 9 The One Touch calibration will assume one of these t...

Страница 10: ...ir 20 9 will result in an incorrect measurement of oxygen concentration 2 2 2 100 Oxygen Calibration 2 6 2 2 Minutes 1 2 3 4 5 6 1 5 Shut off Oxygen supply Remove sensor from port Wait two 2 minutesfo...

Страница 11: ...7 8 2 Minutes Turn flow knob a few turns to flow gas Wait 2 minutes for sensor to equilibrate Press and hold CAL button for three 3 seconds Wait for display to read 100 Remove plug from room air inlet...

Страница 12: ...ettings 4 5 1 2 3 1 2 3 1 2 3 1 2 3 1 3 Adjust flow to new desired flow rate Readjust oxygen concentration to desired setting and wait for analyzer to equilibrate Temporarily disconnect patient circui...

Страница 13: ...ient Interfaces Optiflow large nasal cannula Fisher Paykel OPT544 Optiflow medium nasal cannula Fisher Paykel OPT846 Optiflow Tracheostomy direct connection Fisher Paykel OPT570 Pediatric aerosol mask...

Страница 14: ...Comfort Flo Plus Cannula Large 2412 11 Hudson RCI Comfort Flo Plus Cannula Medium 2412 12 Hudson RCI Comfort Flo Plus Cannula Small 2412 13 F P Direct Connect OPT570 F P Large Nasal Cannula OPT846 F P...

Страница 15: ...sensor to equilibrate to a new ambient temperature ATTENTION CAL Err St may result from a sensor that has not reached thermal equilibrium 4 0 CALIBRATION ERRORS AND ERROR CODES The analyzer has a sel...

Страница 16: ...THE BATTERIES 1 2 3 4 5 6 7 8 9 Turn off analyzer Use screwdriver to remove four 4 screws in back of device Pull front cover off away from unit Turn on analyzer to verify power Reattach battery pack t...

Страница 17: ...s needed The full MaxVenturi device should be overhauled and serviced a minimum of every 4 years with R211P30 001 and R211P30 002 to replace all O rings valves and knobs 7 2 Device Cleaning When clean...

Страница 18: ...s 2 x 1 5 Volts Battery Life approximately 5000 hours with continuous use Low Battery Indication BAT icon displayed on LCD Sensor Type Maxtec MAX 250 series galvanic fuel cell Expected Sensor Life 1 5...

Страница 19: ...d Operating Instructions R125P03 002 MAX 250E Oxygen Sensor RP34P02 Air inlet filter 9 2 Standard Replacement Parts and Accessories Part Number Item R127P35 15 coiled polyurethane oxygen hose with DIS...

Страница 20: ...l knob completely closed Possible Cause It is very likely the device needs to be calibrated Also check to make sure the humidifier and patient delivery disposables are the correct size and installed c...