Maxtec Handi+ N2 Скачать руководство пользователя страница 2

 

2

 

WWW.MAXTEC.COM  •  (800) 748-5355

CLASSIFICATION:

Protection against electric shock .................................................... Internally powered equipment

Protection against water ...............................................................................................................IPX4

Mode of Operation .............................................................................................................Continuous

Flammable anesthetic mixture .....................................................Not suitable for use in presence

of a flammable anesthetic mixture.

Product Disposal Instructions: 

The sensor, batteries, and circuit board are not suitable for regular trash dis-
posal. Return sensor to Maxtec for proper disposal or dispose according to local 
guidelines. Follow local guidelines for disposal of other components.

WARRANTY

The N2 analyzer is designed for oxygen delivery equipment and systems. Under normal 

operating conditions, we warrant the N2 analyzer to be free from defects of workmanship or 

materials for a period of 2- years from the date of shipment provided that the unit is properly 

operated and maintained in accordance with our operating instructions. Based on our product 

evaluation our sole obligation under the foregoing warranty is limited to making replace-

ments, repairs, or issuing credit for equipment found to be defective. This warranty extends 

only to the buyer purchasing the equipment directly from us or through our designated dis-

tributors and agents as new equipment. Our warrants the oxygen sensor in the N2 analyzer 

to be free from defects in material and workmanship for a period of 2-years from the date of 

shipment in a N2 analyzer. Should a sensor fail prematurely, the replacement sensor is war-

ranted for the remainder of the original sensor warranty period. Routine maintenance items, 

such as batteries, are excluded from warranty. We and any other subsidiaries shall not be liable 

to the purchaser or other persons for incidental or consequential damages or equipment that 

has been subject to abuse, misuse, misapplication, alteration, negligence or accident. 

THESE 

WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, 

INCLUDING WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

WARNINGS 

Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.

 

Never allow an excess length of tubing, or any accessory near a person’s head or neck, which 

may result in strangulation.

 

Before use, all individuals who will be using the N2 Analyzer must become thoroughly famil-

iar with the information contained in this Operation Guide. 

 

Strict adherence to the operating instructions is necessary for safe, effective product perfor-

mance. This product will perform only as designed if installed and operated in accordance 

with the manufacturer’s operating instructions.

 

Use only genuine accessories and replacement parts. Failure to do so may seriously impair 

the analyzer’s performance.

 

Repair or alteration of the N2 Analyzer beyond the scope of the maintenance instructions, 

or by anyone other than an authorized service person, could cause the product to fail to 

perform as designed.

 

Calibrate the N2 analyzer weekly when in operation, or if environmental conditions change 

significantly. (ie, Elevation, Temperature, Pressure, Humidity — refer to “Factors Influencing 

Accurate Readings”).

 

Use of the N2 analyzer near devices that generate electrical fields may cause erratic readings

 

Ensure proper tire inflation pressure after use, if required.

 

The oxygen sensor is a sealed device containing a mild acid electrolyte, lead (Pb), and lead 

acetate. Lead and lead acetate are hazardous waste constituents and should be disposed of 

properly, or returned for proper disposal or recovery.

 

Dropping the device can adversely affect its performance.

 DO NOT 

immerse the device in any cleaning solution, autoclave or expose the sensor to 

high temperatures (> 70°C).

 DO NOT 

over pressurize the sensor. Doing so may destroy the sensor and void the war-

ranty. To avoid over pressurization only allow 3 psi (or 2 liters per minute) of gas to come 

in contact with the sensor membrane.

CAUTION 

Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate 
injury and property damage.

 

Read the manual in its entirety before attempting use.

 

Always use protective eyewear and observe proper safety procedures when working with 

pressurized gases.

 

Dispose of the N2 properly when it has expired.

 

Ensure the protective freshness seal has been removed from the sensing port before use.

 

Ensure the N2 has been properly calibrated before use.

 

If the N2 display goes blank immediately after the on button is pushed, or the N2 will not 

properly calibrate, the unit has expired.

 DO NOT 

use, dispose of properly.

 

The display is not valid when in Over Range Mode. Recalibrate the N2 and observe the proper 

operating procedure. 

 

Never immerse the N2 or expose it to high humidity or moisture. It is not watertight.

 

Never expose the N2 to high temperatures.

SYMBOL GUIDE

The following symbols and safety labels are found on the Handi+:

ON/OFF Key

CAL (Calibration Key

Do not throw away. Follow 
local guidelines for disposal

Attention, consult 
accompanying documents

Serial Number

Catalog Number

Corrosive

Lot code/Batch code

DO NOT

Meets ETL standards

Contains Lead

Manufacturer

Caution

Warning

IPX4

Ingress Protection Rating

Maxtec
2305 South 1070 West
Salt Lake City, Utah 84119
USA

phone: (800) 748.5355
fax: (801) 973.6090
email: [email protected]
web: www.maxtec.com

NOTE:

 The latest edition of this operating manual can be downloaded from our website at 

www.maxtec.com