OPERATION
4 - 2
DSE-23
www.maxcess.eu
MI 1092 1
Calibrating the sensor
inputs of the web guide
controller
All settings required for operating the sensor are performed
with the web guide controller.
Information on this topic is available in the Operating
Instructions for the web guide controller.
Before the sensor is used for working with transparent material,
the sensor must always be calibrated to the contrast with the
web guide controller.
LEDs in normal
operation
– Field of view completely uncovered:
The outer LED flickers. The two inner LEDs are off.
Example A
– Field of view partially or completely covered by the material:
The LED on which the edge of the material web is located
flickers. The LED(s) of the covered area(s) is/are lit.
Example B
1 Material web
Figure 4.1: DSE-23 sensor
flickers
off
off
lit
off
flickers
1
1
Example A
Example B