WARNING

 

 

Use only if package is received unopened and package and contents 

are undamaged. If damage is noted, contact the shipper for replacement or 
repair. 

 

Prior to using any MAXAIR System or component, be sure to be familiar with 
the systems’ NIOSH approved configuration, and read and understand the 
User Instructions. 

 

Failure to follow the User Instructions may be hazardous to the user’s 
health. 

 

The institution using this product in any application is responsible for deter-
mining the appropriateness of this equipment relative to regulatory require-
ments. Bio-Medical Devices Intl, Inc does not recommend the appropriate 
systems for a particular institution or facility. 

 

DO NOT use if any component is damaged.  If any components are damaged 
or contaminated and therefore unfit for safe and effective use, they should 
be replaced immediately. 

 

Use only MAXAIR Systems/ NIOSH approved compatible components.   

 

This product contains dangerous voltages. DO NOT attempt to open the 
product cover. This product contains no serviceable parts. DO NOT attempt 
to modify any component.   

 

If you need more information, contact your BMDI sales representative or call  
BMDI customer service at 1-800-443-3842. 

 

CAUTIONS 

 

The chargers are only to be used with MAXAIR Systems Lithium-Ion batteries.  

 

The chargers have internal fuses which blow if a fault occurs in the charger. In 
addition, they are equipped with a fuse switch, which cuts off the unit in the case 
of reverse polarity connection to the battery.  

 

Do not leave any battery connected to a powered-on charger for longer than one 
month. Refer to MAXAIR Systems User Instructions Manual (UIM) for further 
details regarding intermittent use and storage of batteries. 
 
The chargers have a plastic casing - avoid charger contact with oils, grease, etc. 
as most plastics can be broken down by chemicals and solvents. 
 

SYMBOLS DEFINITION 

 

Read these instructions before using the charger.  

 

This product is designed for indoor use only and should not come into con-
tact with water or dust. To prevent overheating the product should not be 
covered during use.  

 

The mains socket should be easily accessible. In the event of operational 
error, the plug should be immediately removed from the socket.  

 

The charger is double-insulated (Class II). 

 

The product label “EN60601-1” indicates it complies with the requirements 
of  electromedical equipment and can be used in hospital environments. The 
product should not be used in the vicinity of flammable anesthesia gases.  

 

UL Recognized Component Marks for Canada and the United States.  

CHARGING INSTRUCTIONS 

 

1. Connect the charger (single and/or 6-Gang) to an appropriate grounded 
outlet power source (100-240 VAC, 50-60 Hz ) before connecting battery(ies). 
The Charger green status LED should turn on. 

 

2. Connect the battery(ies) to the charger(s). The Charger status LED should 
change from green to red to indicate the battery is charging. 

 

3. When charging is complete, the Charger LED should change to green. 
Disconnect battery from charger. The battery is ready for use. 
 

LED’s INDICATE FOLLOWING CHARGE STATUS 

 

Fast charge—LED indicator is red

 

 

The charger is in constant current mode. 

 

Charge current is maximum. 

 

Final charge—LED indicator is red 

 

The charger is in constant voltage mode. 

 

Charge current is less than maximum. 

 

The battery is normally 80-95% charged. 

 

The charger stays in this mode until the charge  

current decreases to charge termination level.  

 

Charge completed—LED indicator is green 

 

Charging has stopped.  

 

Charge current is zero.  

 

LED indicator changes to green. 

 

GENERAL SPECIFICATIONS 

ORDER NO:2600-01  ELECTRICAL OUTPUT 

ELECTRICAL INPUT 

PART NO: 01432089  VOLTAGE: UP TO 16.8V 

VOLTAGE: 100-240 VAC, 
50-60Hz max 

 

CURRENT: UP TO 0.9A 

CURRENT: 0.3A 

Intended Use 

 

The MAXAIR Systems 6-Gang UL/CE Charger (2601-06) and  
Charger Bracket (2601-06B) are intended for use only when charging  
MAXAIR Systems’ Lithium-Ion Batteries. 

 

The 6-Gang Charger includes six 2600-01 UL/CE Chargers and is  
capable of charging up to six MAXAIR Systems’ Lithium-Ion batteries 
simultaneously. 

 

The Charger Bracket requires from one to six MAXAIR 2600-01 
Chargers to be assembled (instructions below) with it to allow charging 
of from one to six MAXAIR Lithium-Ion batteries. 

 

Description 

 

The 6-Gang Charger is fully assembled and ready to charge one to six  
MAXAIR Lithium-Ion batteries by plugging it into a properly grounded  
110 VAC mains line. 

 

The 6-Gang Charger kit requires the addition of one to six MAXAIR 
2600-01 Chargers, and then is similar to the 6-Gang Charger. 

P/N 01423158 REV A 

2601-06 
6-Gang  
Charger 
(UL/CE) 

2601-06B 
6-Gang  
Charger Bracket 

(for 2600-01 Chargers) 

NOTE:  
1 connected  
battery  
(NOT INCLUDED) 
shown in photo. 