5.6.2 Alarm current output
Sets output current value during fault condition.
Alarm current selection: Output current value when measurement is not valid because of no
reflection echo loss or other reasons.
Selection =
「
Hold
」
「
Sel.val.
」
「
Max
」
「
Min
」
・
Hold:
Output remains at previous measured result during alarm.
・
Selected value (Sel.val.): Alarm output current is set to value entered in “5.Sel. alarm
current val.”
・
Max: Alarm output current is set to value selected in “3. Max current sel.”
・
Min: Alarm output current is set to value selected in “4.Min current sel.”
Max current sel.: Selects current output value when
「
Max
」
is selected in 2. Alarm output sel.
Selection range =
「
20mA
」
「
20.5mA
」
「
22mA
」
Min current sel.: Selects current output value when
「
Min
」
is selected in 2. Alarm output sel.
Selection range =
「
4mA
」
「
<3.6mA
」
「
3.8mA
」
(Microwave Level Meter over Ver2.0)
Selection range =
「
4mA
」
「
3.6mA
」
「
3.8mA
」
(Except for the above)
Sel. alarm current val: Sets current output value when
「
Sel.val.
」
is selected in 2.Alarm output sel
Selection range = 3.55mA to 22.0mA (Microwave Level Meter over Ver2.0)
Selection range = 3.60mA to 22.0mA (Except for the above)
※
You can not select several options for alarm current output.
12.345m
Distance
Menu
0
1. Identification setting
2. Basic setting
3. Sensor adjustment
4. Graph display
5. Self test
6. Reset
▼
Menu
0
1. Identification setting
2. Basic setting
3. Sensor adjustment
4. Graph display
5. Self test
6. Reset
▼
Menu view will be
displayed.
Push
『+』
twice shift
cursor down to
Sensor adjustment
menu.
Push
『Ent』
to
enter Sensor
adjustment menu.
Sensor adjustment
3
1.Current output set.
2.Meas. adjustment
Current output set. 3.1
1.Current output sel.
= 4-20mA
2.Alarm current sel.
= Hold
3.Max current sel.
= 20mA
▼
Sensor adjustment
menu will be
displayed.
Push
『Ent』
to enter
Current output set.
Ent
× 1 click
Ent
× 1 click
Ent
× 1 click
Ent
× 1 click
+
× 2 click
Current output set. 3.1
1.Current output sel.
= 4-20mA
2.Alarm current sel.
= Hold
3.Max current sel.
= 20mA
▼
Push
『Ent』
to enter
Alarm current sel.
Current output set. 3.1
1.Current output sel.
= 4-20mA
2.Alarm current sel.
=
Hold
3.Max current sel.
= 20mA
▼
+
× several times
Push
『+』
to change
value to Sel.val.
Push
『Ent』
set to
selected value.
Current output set. 3.1
1.Current output sel.
= 4-20mA
2.Alarm current sel.
= Sel.val.
3.Max current sel.
= 20mA
▼
Push
『+』
3 times shift to
Sel. alarm current val.
Cont’d on next page
+
× 3 click
Ent
× 1 click
+
【
Measurement view
】
Push
『Ent』
to enter
main menu.
Current output setting
view will be displayed.
Example: Set alarm
current to 21mA.
Push
『+』
shift to Alarm
current sel.