6 Operating by Analog Remote Control
49
HARS series
6-3 Using Interlock
You can turn the output Force ON/OFF using Output Switch. This function can be implemented according
to the following method.
Set the interlock switch to ON/OFF (short = output enabled, open = output OFF*).
If the switch is open, the touch panel display will show an "INTERLOCK" error message.
When shorted, output is enabled by the OUTPUT ON/OFF switch or External Switch.
*When the interlock function is disabled, the interlock protection function is not available.
TTL LOW
=
OUTPUT:ON enable
TTL HIGH
=
OUTPUT:OFF
or
External relay SHORT
=
OUTPUT:ON enable
External relay OPEN
=
OUTPUT:OFF
13
25
14
1
TB1
TTL signal
External relay
External connection
+
-
8
17
13
25
14
1
Internal circuit
of unit
470Ω