4 Operation
47
DOEF series
Matsusada Precision
The following table shows the types of faults and warning and display contents
Display
State
FAULT (E24) OCP !
Fault: Cutoff protection is activated due to overcurrent.
FAULT (E22) OVP !
Fault: Cutoff protection is activated due to overvoltage.
FAULT (E23) OPP !
Fault: Sink power protection is activated.
OVER POWER !
Warning: Sink power exceeds the set value.
INTERLOCK !
Fault: Interlock is activated.
FAULT (E02) OT !
Fault: Over temperature protection is activated.
4-2-4 Fault Release
Press the OUPUT switch to release the fault state. In addition, sending a LS signal or communication
command to turn on the output is also possible to release the fault state.
When over-temperature protection has been activated:
Turn off the unit and wait for 30 minutes or more to turn it on again.
If the protection is activated continuously, it may be something wrong with it, please contact
us.