6
9.
Remove the flushcap (
4
) by turning the knurled ring (
5
) counter-clockwise until
the flushcap can be removed.
a.
Note:
As long as a purifier cartridge is not installed, no gas will
pass through the baseplate.
10.
Install a new purifier cartridge on the baseplate. See SECTION VIII,
Installing a Purifier Cartridge, for more detail.
11.
Flush the whole system for 15 minutes with pure carrier gas, at a recommended
pressure of 90 psig.
12.
Tighten the outgoing gas line at the Outlet (
2
) rear fitting.
The analytical system is now ready to be used. Argon or Nitrogen peaks in the
chromatogram should be negligible.
➢
NOTE:
Always check for possible leaks in the system before operation.