background image

 

 

 

 

9.

 

Remove the flushcap (

4

) by turning the knurled ring (

5

) counter-clockwise until 

the flushcap can be removed. 

 

a.

 

Note: 

  As long as a purifier cartridge is not installed, no gas will 

pass through the baseplate.

 

 

10.

 

Install a new purifier cartridge on the baseplate.    See SECTION VIII, 
Installing a Purifier Cartridge, for more detail.   

11.

 

Flush the whole system for 15 minutes with pure carrier gas, at a recommended 
pressure of 90 psig. 

12.

 

Tighten the outgoing gas line at the Outlet (

2

) rear fitting. 

 
The analytical system is now ready to be used.    Argon or Nitrogen peaks in the 
chromatogram should be negligible. 
 
 

 

NOTE: 

  Always check for possible leaks in the system before operation. 

 
 

 

Содержание PUR-Gas MCTG-0050

Страница 1: ...MINT 0248 rev B PUR Gas Purifier System Instructions READ AND COMPLY WITH THESE INSTRUCTIONS BEFORE INSTALLING OPERATING OR SERVICING ...

Страница 2: ......

Страница 3: ...R RESPONSIBILITY 3 IV SAFETY PRECAUTIONS 4 V GENERAL DESCRIPTION 4 VI INSTALLATION OF A BASEPLATE 5 VII STANDARD PURIFIER CARTRIDGE BASEPLATE COMBINATION 7 VIII INSTALLING A PURIFIER CARTRIDGE 8 IX REPLACEMENT OF A PURIFIER CARTRIDGE 8 X ORDERING INFORMATION 11 ...

Страница 4: ......

Страница 5: ...uct will be received by MATHESON without indication of gas service and without proper return material authorization provided by the warranty administrator If advised by the Warranty Administrator to return the product to MATHESON prepare the product for shipment and write in large lettering the RMA Number assigned by the Warranty Administrator on the outside of the box Also if required by the Warr...

Страница 6: ...n operated and maintained in accordance with written instructions provided by MATHESON or has been altered by anyone other than MATHESON or if the equipment has been subject to abuse misuse negligence or accident MATHESON s sole and exclusive obligation and the Buyer s sole and exclusive remedy under the above warranties is limited to repairing or replacing free of charge at MATHESON s sole discre...

Страница 7: ...replacement become necessary MATHESON recommends that a telephonic or written request for service advice be made to the MATHESON Equipment Engineering Group in Montgomeryville Pennsylvania or to the nearest MATHESON branch location This equipment or any of its parts should not be altered without the prior written approval of MATHESON Equipment Engineering Group The user of this equipment shall hav...

Страница 8: ...artridge check for any cracks on the inner glass tube after installation 7 NEVER REMOVE the white plastic mounting plugs located on the outside housing near the bottom of the cartridge if they are missing the cartridge should NOT be installed and should be returned to MATHESON 8 NEVER replace the baseplate fittings if new fittings are required contact MATHESON for technical assistance 9 Use only o...

Страница 9: ...ate metal tubing cutter 3 Connect the gas Inlet 1 and Outlet 2 as indicated at the rear of the baseplate keeping the connections loose 4 Turn on the gas source and flush the line for at least 10 minutes with pure carrier gas at 8 psig minimum 5 Tighten the incoming gas line at the Inlet 1 rear fitting 6 Ensure that the flushcap 4 is securely tightened to the baseplate push the cap down during moun...

Страница 10: ...artridge on the baseplate See SECTION VIII Installing a Purifier Cartridge for more detail 11 Flush the whole system for 15 minutes with pure carrier gas at a recommended pressure of 90 psig 12 Tighten the outgoing gas line at the Outlet 2 rear fitting The analytical system is now ready to be used Argon or Nitrogen peaks in the chromatogram should be negligible NOTE Always check for possible leaks...

Страница 11: ...fication System for GC FID High Capacity Triple 2x Combi purifier on 3 position baseplate 2 Purification System for Fuel Gas FID 2x Combi purifier on 2 position baseplate 3 Purification System for Carrier Gas with MS ECD NPD Triple purifier on 1 position baseplate ...

Страница 12: ...ed on the baseplate the Teflon seals will be punctured It is necessary to keep the purifier cartridge in position to prevent air from entering the system 2 Regularly check both the big and small Viton O rings on the baseplate for hair cracks The slightest leak allows moisture oxygen and hydrocarbons to enter the system and contaminate the gas 3 All MATHESON PUR Gas purifier cartridges are packed u...

Страница 13: ... Cartridge Label 1 Type of purifier cartridge 2 MATHESON part number for purifier cartridge 3 Indicator color before saturation 4 Indicator color after saturation 5 Installation date to be filled in by user 6 Month Year of production 1 2 4 5 6 3 ...

Страница 14: ...en 1000 Hydrocarbon Not specified Combi 3 6 Not specified Triple 1 8 500 Not specified Note The data given in the table strongly depends on the level of gas purity prior to entering the purifier Note Replacement of the purifier cartridge is recommended at least every six months regardless of the saturation level ...

Страница 15: ...sition Baseplate with 1 4 SS Fittings MBAS 0027 2 Position Baseplate with 1 4 SS Fittings MBAS 0028 3 Position Baseplate with 1 4 SS Fittings MBAS 0029 2 Position LC MS Baseplate with 1 4 SS Fittings MCTG 0050 Moisture Cartridge MCTG 0051 Oxygen Cartridge MCTG 0052 Hydrocarbon Cartridge MCTG 0053 Triple Cartridge O2 H2O HC MCTG 0054 Combi Cartridge H2O HC MCTG 0055 LC MS Charcoal Cartridge Set 2 M...

Страница 16: ...nc cannot guarantee the accuracy of this material nor can it accept responsibility for errors or omissions This document is intended to serve as a general orientation and cannot be relied upon for a specific operation No warranties of any nature are extended by the information contained in these copyrighted materials All names products and services mentioned herein are the trademarks or registered...

Отзывы: