background image

TROUBLESHOOTING

Troubleshooting

SYMPTOM

CAUSE

REMEDY

Defective line

1. Check that unit is plugged into a

cord or fuse.

live line.

2. Check line cord for continuity and 

replace if defective.

3. Check fuse and replace if defective.

Defective

1. Place the FWD-OFF-REV switch

Motor or 

in the OFF position.

Controller.

2. Test motor and controller.

a). Check the drive motor cable

connection to the controller.

b). Replace motor, with similar 

unit. Reconnect the drive 
motor to the controller. Retest.

c). Replace controller with similar

unit. Reconnect the drive 
motor to the controller. Retest.

Motor does not
rotate, when
switched to FWD or
REV. POWER indica-
tor does not glow.

Motor does not
rotate, when
switched to FWD or
REV. Power indica-
tor glows. Speed
control setting is
greater than 0.

SPECIFICATIONS

Operating Temperature:

0˚ to 40˚C

Storage Temperature:

-45˚ to 65˚C

Chemical Resistance:

Exposed material is
paint, plastic, aluminum,
and vinyl

Line Voltage Limits:

7552-70

100–130 V, 50-60 Hz

7552-75

190–260 V, 50-60 Hz

Speed Range:

6 to 600 rpm

Typical Power:

Motor:

0.075 kW (0.1 hp)

Controller:

6 watts, max. dissipation

Fuses:

115V Units:

3 A

230V Units:

1.5 A

Torque:

13 kg-cm (180 oz-in)

Speed Regulation:

±5% Line Regulation
±10% Load Regulation
±10% Warm-Up Drift

Controller Enclosure Rating:

IP56

Motor Enclosure Rating:

IP56

Humidity: (non-condensing)

0% to 90% 

Display:

RED Indicator

Dimensions: (L x W x H)

Motor:

280mm x 102mm x 127mm

(11 in x 4 in x 5 in)

Controller:

229mm x 203mm x 152mm
(9 in x 8 in x 6 in)

Weight:

Motor:

4.1 kg (9 pounds)

Controller:

2.3 kg (5 pounds)

6

7