5. Meter maintenance

This section provides basic maintenance information, 
including instructions for replacement of fuse and battery. 
Do not try to repair the meter unless you are an experienced 
maintenance person with the relevant calibration, performance 
testing and maintenance data

5.1 General maintenance

To avoid electrical shock or damage to the meter, don’t wet the 
inner surfaces of the meter. Before opening shell or battery 
cover, you should remove the connecting cable between the 
test probe and the input signal. 

Regularly clean the meter shell with damp cloth and a small 
amount of detergent. Do not use abrasives or chemical solvents. 
If you make input socket dirty or wet, it may affect the readings. 

To clean input socket: 

Turn off the meter, and pull out all test probes from the input 
socket. 
Remove all dirt from the socket. 
Apply detergent or lubricant (such as WD-40) to a new cotton 
ball. 
Clean each socket with a cotton ball. Lubricant can prevent 
contamination related with moisture on the socket. 

5.2 Replace battery and fuse

To avoid electrical shock or personal injury, before opening the 
battery cover to replace battery, you should turn the meter off 
and make sure that the test probe is disconnected from the 
measurement circuit. 

5.3 Replacing Test Leads

WARNING: 

WARNING: 

WARNING: 

Use meet EN 61010-031 standard, rated CAT III 600V,
10A

 

or better test leads.

Please follow below steps to replace battery or fuse: 
1. Turn off the power supply of the meter. 
2. Pull out all test probes from the input socket. 
3. Loosen two screws on the fixed battery cover with screwdriver. 
4. Remove the battery cover. 
5. Remove the old battery or damaged fuse. 
6. Replace with a new battery with 9V (6F22) or a new fuse. 
7. Replace the battery cover and tighten the screws.

To avoid wrong readings, electric shock or personal injury, when
 “

  

”appears on the meter display, replace the battery 

immediately.  
Use only a fuse with specified amperage, fusing rated value, 
voltage rated value and fusing speed (F1:FF 400mA H 600V, 
F2:FF 10A H 600V) 
    

If insulation on leads is damaged, replace it.

1. The meter can be cleaned with a soft cloth to remove any 
oil, grease or grime.
2. Do not use liquid solvent or detergent. 

5.4 Cleaning and Decontamination

6. Accessories

• 
• 
• 
• 

Test Leads                                         1set 
Package                                             1pcs
9V  Battery                                         1pcs
User's Manual                                    1pcs

Содержание MY70

Страница 1: ...MY70 74 74B OPERATION MANUAL DIGITAL MULTIMETER CAT III 600V ...

Страница 2: ...iode 11 3 2 4 Audible continuity testing 12 3 2 5 Measuring capacitance MY74 MY74B 12 3 2 6 Transistor measuring 3 2 7 Measuring frequency MY74 MY74B 3 2 8 Measuring temperature MY74 MY74B 3 2 9 Measuring current 4 Technical indicators 4 1 Comprehensive indicators 4 2 Accuracy indicators 4 2 1 DC voltage 4 2 3 Frequency only for MY74 MY74B 4 2 4 Resistance 4 2 5 Diode 4 2 6 Audible continuity 19 4...

Страница 3: ...meter When the meter is connected to the circuit being measured do not touch the unused input end When the voltage to be tested exceeds 60Vdc or 30Vac effective value please operate carefully to prevent electric shock When you measure with test probe first connect the common testing end of black test probe to the common testing end of circuit to be tested then connect red test leads to the test ci...

Страница 4: ...ircuits being tested The TV filter should be used to weaken these pulses to avoid the meter damage This meter uses a 9V 6F22 battery The battery should be properly installed in the meter s battery compartment When the battery indicator appears the battery should be replaced immediately Low battery will cause meter reading errors and possibly result in electric shock or personal injury When you mak...

Страница 5: ...nomena on the meter the meter should be immediately turned off and repaired Ensure that it can not be used before passing inspection When the meter is not used for a long time please remove the battery and avoid storing it in a high temperature and humidity environment 1 2 Input protection measures When making voltage measurements not including 200mV grade the maximum input voltage is 600V DC or 6...

Страница 6: ... the battery should be replaced immediately Symbols Indication Negative input polarity indication Keep the current measurement value HOLD Table 2 Input Socket COM mA 2 2 LCD display See Table 1 for information about the display Fig 1 Display HV High voltage symbol in AC600V or DC600V Positive input end of voltage resistance frequency diode buzzer measurement and temperature test connected with the...

Страница 7: ...e 1 Rotate the switch to the appropriate position 2 Respectively connect black and red test probe to COM input socket and V input socket 3 Measure the voltage of circuit to be tested with other two ends of test probes Connected with the circuit to be tested in parallel 4 Read the measuring voltage value from LCD display When measuring DC voltage the display will simultaneously show the voltage pol...

Страница 8: ...diode should still produce a forward voltage drop of 0 5V to 0 8V but the reverse bias voltage will vary depending on resistance values of other channels between two test probes 3 2 4 Audible continuity testing To avoid damaging meter or device to be measured before testing buzzer continuity cut off all circuits being tested and discharge all high voltage capacitors To make continuity testing 1 Ro...

Страница 9: ...h electric quantity higher than 60V DC or 24V AC RMS to prevent electrical shock Don t measure the temperature in a microwave oven to prevent fire or meter damage 3 2 9 Measuring current When the ground voltage of open circuit voltage exceeds 250V do not try to make current measurements on the circuit If the fuse is blown when making a measurement you may damage the meter or injure yourself To avo...

Страница 10: ...g range 6 Cut off the power supply of circuit to be tested Discharge all high voltage capacitors Remove the test probe of meter and restore the circuit to its original condition 4 Technical indicators 4 1 Comprehensive indicators Operating environment and condition 600V CAT III pollution grade 2 Elevation 2000 m Environment temperature and humidity O 0 40 C 80 RH 10 C it is not to be considered St...

Страница 11: ... or 600V ac RMS 250Vdc or ac RMS with the measuring range of 200mV Frequency response 40Hz 400Hz sine wave RMS average response The frequency response is 200Hz for 600V 4 2 3 Frequency only for MY74 MY74B Measuring range Resolution Accuracy 20kHz 10Hz 2 0 of reading 5digits Input voltage range 200mV 10V ac RMS Overload protection 250V dc or 250V ac RMS 4 2 4 Resistance Measuring range Resolution A...

Страница 12: ...eading 1digits 20μA for MY70 MY74B 200μA only for MY70 20mA 200mA 2mA for MY70 MY74 10A 0 01μA 0 1μA 1μA 10μA 0 1mA 10mA 1 5 of reading 1digits 2 0 of reading 5digits Overload protection fuse with measuring range of mA FF400mA H 600V fuse with measuring range of 10A FF10A H 600V Maximum input current mA grade 200mA DC or AC RMS 10A grade 10A DC or AC RMS When measured current is greater than 10A c...

Страница 13: ... avoid electrical shock or personal injury before opening the battery cover to replace battery you should turn the meter off and make sure that the test probe is disconnected from the measurement circuit 5 3 Replacing Test Leads WARNING WARNING WARNING Use meet EN 61010 031 standard rated CAT III 600V 10A or better test leads Please follow below steps to replace battery or fuse 1 Turn off the powe...

Страница 14: ...00 05 4176 ...

Отзывы: