5. Maintenance

5.1

 

General Maintenance

This section provides basic information on maintaining 
the meter, such as replacing fuses and the battery. Only 
experienced and authorized personnel should make 
repairs to the meter.

Warning

To avoid injury or damage to the meter,do not allow
moisture inside the case and remove test leads 
before opening battery cover.

• Use a damp cloth to regularly clean the outside of the 
  meter.  Do not use abrasives or chemical solvents.Dirty 
  or damp input jack can adversely affect readings.
• To clean input jacks, follow the following steps:
1.Turn off the instrument and remove the test leads.
2.Clear any dirt or other particles on the input jacks.
3.Use a cotton ball/swab with a lubricant (i.e. WD-40) to 
   clean off the contacts of the input jacks.
4.Use a separate cotton ball/swab for each jack to 
   prevent cross-contamination.

Note:

• Even if the indicator doesn’t go off, voltage may still 
  exist.Do not use NCV detection as the sole 
  determination of a lack of voltage present in a 
  conductor.Detection could be impaired by socket 
  design, insulation, etc.
• When measuring voltage through input jacks, NCV 
  indicator may light up due to induced voltage.
• External environmental interference could cause false 
  NCV indications.

5.2

 

Replacing the Battery and Fuses

5.3

 

Replacing the Probe

1. Under normal conditions, it is unnecessary to replace the fuse. 
    Don't replace it until the probes are unplugged and the power is 
    shut down. Take out the two screws of the rear cover to remove 
    the housing.
2. The specification of the fuse is:
    F1 F 10A H 600V, F2 F 400mA H 600V
    The replacement should be of the same specification.
3. The battery for this multimeter is 9V NEDA 1604 or 6F22. 
    The replacement should be of the same specification.
4. Don't put the instrument into use until the rear cover is screwed 
    after replacing battery or fuse.

To avoid electric shock, make sure the probes are disconnected 
from the measured circuit before removing the rear cover.
Make sure the rear cover is tightly screwed before using the 
instrument.

       Warning

If insulation on probe is damaged, replace it.

Use meet EN 61010-031 standard, rated CAT III 600V, 10A or better probe.

       WARNING

6. Accessories

• 
• 
• 
• 

Test Leads                                          1set 
Package                                             1pcs
9V  Battery                                         1pcs
User's Manual                                    1pcs

Содержание MS8238B

Страница 1: ...gital Multimeter CAT III 600V MS8238B DIGITAL MULTIMETER HOLD NCV 200 600 200m 200m 200 2k 20k 200k 2M 2 20 200 600 200m 9V 1 5V 12V A V V A 600V CAT III mAV COM MAX 30 sec every 15 min MAX FUSED 10A...

Страница 2: ...mbols on the meter are to advise of potential dangerous situations Caution is required when measuring close to the mete s safety limits Never exceed the protection limit values indicated in the specif...

Страница 3: ...ower and check fused before connect the leads when measuring current Turn circuit power on after making connection Never use the meter unless the back cover is in place and fastened securely When the...

Страница 4: ...ment display 4 Polarity indicator 3 4 1 2 8 1 2 7 3 4 5 6 6 Input jack all functions except current greater than 200mA 7 Non contact voltage NCV button 8 Non contact voltage NCV indicator MS8238B DIGI...

Страница 5: ...current range Move the rotary switch to a higher range Set the rotary switch to the appropriate AC or DC Warning Do not measure voltages higher than 600V DC or AC rms to perevent damage to the meter o...

Страница 6: ...red test lead to the input jack and the black lead to the COM jack Connect the leads to the circuit under test If the measured resistance is less than 50 the buzzer will sound Warning To prevent inju...

Страница 7: ...LCD display Overload Indication Display shows OL Low Battery Indication 4 2 Technical Specifications Accuracy of reading digits at 18 C 28 C with a relative humidity of 80 guaranteed for a period of...

Страница 8: ...k FF10A H 600V fuse Max input current mA jack 200mA DC 10A jack 10A DC When measuring current exceeding 2A do not measure for longer than 2 minutes continuously Wait 10 minutes to continue measurement...

Страница 9: ...ection could be impaired by socket design insulation etc When measuring voltage through input jacks NCV indicator may light up due to induced voltage External environmental interference could cause fa...

Страница 10: ...R 00 05 1049...

Отзывы: