4.8 Frequency
Range
Accuracy
Resolution
2kHz
±1.5% of rdg ±5digits
1Hz
5. Accessories
5.1 Supplied with the multimeter
Test Leads
Battery
HYTL- 266
Operating Manual
9V NEDA 1604 or 6F22
HYS004249A1
5.2 Optional accessory
“K”type thermocouple
HYTP - 010
INSULATION TESTER UNIT
261
6. Battery replacement
If the sign“BAT”appears on the LCD display, it indicates
that battery should be replaced. Remove the battery cover
of case. Replace the exhausted battery with a new one.
Warning
Before attempting to open the battery cover, be sure that
test leads have been disconnected from measurement
circuits to avoid electric shock hazard.
4.5 DC Voltage
Input Impedance: >9M
Ω
Over protection:
250Vrms AC for 200mV range
1000V DC or 750V AC for other range
4.6 Resistance
Overload protection: 250V DC or 250Vrms AC on all ranges.
Open circuit voltage: 700mV
Range
Accuracy
Resolution
200mV
±5.0% of rdg ±1digit
0.1mV
2V
±5.0% of rdg ±1digit
1mV
20V
200V
1000V
±5.0% of rdg ±1digit
±5.0% of rdg ±1digit
±0.8% of rdg ±2digits
10mV
0.1V
1V
Range
Accuracy
Resolution
200
Ω
±1.0% of rdg ±3digits
0.1
Ω
2k
Ω
±1.0% of rdg ±1digit
1
Ω
20k
Ω
200k
Ω
2M
Ω
±1.0% of rdg ±1digit
±1.0% of rdg ±5digits
±1.0% of rdg ±5digits
10
Ω
100
Ω
1k
Ω
4.7 Temperature
Range
Accuracy
Resolution
0°C to 400
32
to 752
°C
(
°F
°F)
400°C to 750
752
to 1382
°C
(
°F
°F)
0°C to
750°C
°F
°F
32
to
1382
1°C
°F
1
±1.0% of rdg ±3digits ±2.0% of rdg ±1digit
Display
Terminals and
earth ground
Measuring Method
Overrange Indication
Polarity indication
Operating Temperature
Storage Environment
Power
Accessories
Low Battery Indication
Optional Accessories
Dimension
Weight
3 1/2 digit LCD,
with automatic polarity indication
1000V dc or 750V rms ac (sine)
“1” Figure only in the display
“-” displayed for negative polarity
0°C to 40°C (32°F to 104°F)
9V alkaline or carbon-zinc battery
(NEDA 1604 )
Operating manual ,set of test leads
“BAT” to left of display
Thermocouple ( K type )
90(W) X 230(D) X 43(H) mm
320g( including battery )
Dual-slope integration A-D converter
-10°C to 50°C (14°F to 122°F)
3.8 Measuring frequency
4. Specifications
Accuracy is specified for a period of one year after
calibration and at 18°C to 28°C (64°F to 82°F) with relative
humidity to 80%.
1. Connect the black test lead to the COM jack and the red
test lead to the V
Ω
jack.
2.Set the rotary switch at Hz position and connect test leads
across the source or load under measurement.
NOTE:
1. Reading is possible at input voltage above 10V rms,
but the accuracy is not guaranteed.
2. In noisy environment, it is preferable to use shield cable
for measuring small signal.
3.3 Measuring voltage
1. Connect the black test lead to the COM jack and the red
test lead to the V
Ω
jack.
2. Set the rotary switch at the desired V or V ~ range
position and connect test leads across the source or
load under measurement. The polarity of the red lead
connection will be indicated along with the voltage value
when making DC voltage measurement.
3. When only the figure“ 1 ”is displayed, it indicates
overrange situation and the higher range has to be
selected.
3.4 Measuring resistance
1. Connect the black test lead to the COM jack and the red
test lead to the V jack.
2. Set the rotary switch at desired position and connect
test leads across the resistor under measurement.
NOTE:
1. If the resistance being measured exceeds the maximum
value of the range selected or the input is not connected,
an overrange indication“ 1 ”will be displayed.
2. When checking in - circuit resistance, be sure the circuit
under test has all power removed and that all capacitors
have been discharged fully.
Ω
Ω
3.6 Continuity test
1. Connect the black test lead to the COM jack and the red
test lead to the V
Ω
jack.
(The polarity of the red lead is positive“ + ”)
2. Set the rotary switch at position and connect test
leads across two points of the circuit under testing.
If continuity exists (i.e., resistance less than about 50
Ω
),
built -in buzzer will sound.
3.7 Measuring temperature
1. Set the rotary switch at °C or °F position and the LCD
display will show the current environment temperature.
2. Insert“ K ”type thermocouple into the temperature
measuring socket on the front panel and contact the
object to be measured with the thermocouple probe.
Read LCD display.
Warning
To avoid electric shock, be sure the thermocouple has been
removed before changing to another function measurement.
3.5 Testing diode
1. Connect the black test lead to the COM jack and the red
test lead to the V
Ω
jack. (The polarity of red lead is“ + ”)
2. Set the rotary switch at position and connect red
lead to the anode, black lead to the cathode of the diode
under testing. The meter will show the approx. forward
voltage of the diode. If the lead connection is reversed,
only figure“ 1 ”displayed.
09
10
11
4.2 AC Current
Range
Accuracy
Resolution
20A
±5.0% of rdg ±5digits
10mA
Frequency Range: 50Hz to 60Hz
Response: Average, Calibrated in rms of sine wave
Overload protection: 1200A within 60 seconds.
Jaw Opening: 2" (5cm
200A
400A
1000A
±2.5% of rdg ±5digits
±2.5% of rdg ±5digits
±3.0% of rdg ±5digits
100mA
1A
1A
Note
>600A
4.3 Insulation test
Range
Accuracy
Resolution
±2.0% of rdg ±2digits
±4.0% of rdg ±2digits
±5.0% of rdg ±2digits
Note
20M
Ω
2000M
Ω
1M
Ω
1k
Ω
>500M
Ω
4.4 AC Voltage
Range
Accuracy
Resolution
200V
±1.0% of rdg ±4digits
0.1V
Input Impedance: >9M on all ranges
Overload protection: 1000V DC or 750V AC on all ranges
Frequency Range: 50Hz to 500Hz
Response: Average, calibrated in rms of sine wave
750V
±1.0% of rdg ±4digits
1V
< 600A
< 500M
Ω
HYS006730
6. Depress the PUSH 500V push-push switch, the 500V on
red LED lamp will light. Clamp meter display reading is
the insulation resistance value. If the reading is below
19M
Ω
, change clamp meter and insulation tester unit to
20M
Ω
range, can be increase the accuracy.
7. If the insulation tester unit is not use, the power switch
must shift to power OFF position, and the test leads must
leave the E. L input connect. That can be increase battery
life and prevent electrical shock hazard.
4.1 General
(With option 500V insulation tester unit)