Masimo Radius T Скачать руководство пользователя страница 2

ENVIRONMENTAL

Storage/Transport Temperature

-20°C to 50°C @ ambient humidity

Operating Temperature

10°C to 40°C @ ambient humidity

Storage/Transport Humidity

10% RH  to 95% RH (non-condensing) @ ambient temperature

Operating Humidity

10% RH to 95% RH (non-condensing) @ ambient temperature

WIRELESS TECHNOLOGY INFORMATION

Type

Bluetooth Low Energy

Data Transmission Rate

Minimum packet rate of 0.0167 Hz (1/60 Hz)

Max. Output Power

(EIRP): 9.9 dBm

Modulation Type

GFSK

Frequency Range

2402–2480 MHz

Antenna Peak Gain

+5.67dBi

FCC ID are as follows:  FCC ID: VKF-RADIUST, IC ID: 7362A- RADIUST

CAUTION:

 In order to maintain Bluetooth connectivity with the host device ensure that Radius T° is within 

specified distance and line of sight of the host device. 

RF Radiation Exposure Statement: 

This equipment has been exempted from FCC RF radiation exposure 

testing and IC RSS 102 RF radiation exposure limits set forth for an uncontrolled environment. 

Note: 

This device complies with part 15 of FCC Rules and Industry Canada’s license-exempt RSSs. Operation is 

subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device 
must accept any interference received, including interference that may cause undesired operation.

Note: 

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant 

to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful 
interference in a residential installation. This equipment generates, uses, and can radiate radio frequency 
energy and, if not installed and used in accordance with the instructions, may cause harmful interference 
to radio communications. However, there is no guarantee that interference will not occur in a particular 
installation. If this equipment does cause harmful interference to radio or television reception, which can be 
determined by turning the equipment off and on, the user is encouraged to try to correct the interference by 
one or more of the following measures:

•  Reorient or relocate the receiving antenna.

•  Increase the separation between the equipment and receiver.

•  Consult the dealer or an experienced radio/TV technician for help.

Note: 

When using Radius T° consideration should be taken to local government frequency allocations and 

technical parameters to minimize the possibility of interference to/from other wireless devices.

RECOMMENDED SEPARATION DISTANCES

RECOMMENDED SEPARATION DISTANCE BET WEEN PORTABLE AND   

MOBILE RF COMMUNIC ATION EQUIPMENT AND THE ME EQUIPMENT

The ME Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The 
customer or the user of the ME Equipment can help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and the ME Equipment as recommended below, 
according to the maximum output power of the communication equipment.

RATED MAXIMUM OUTPUT POWER OF 

TRANSMIT TER (W)

SEPARATION DISTANCE ACCORDING TO FREQUENC Y OF TRANSMIT TER (M)

150 kHz to 80 MHz
d = 1.17*√P

80 MHz to 800 MHz
d = 1.17*√P

800 MHz a 2.5 GHz
d = 2.33*√P

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.17

1.17

2.33

10

3.7

3.7

7.37

100

11.7

11.7

23.3

For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can 
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of 
the transmitter in watts (W) according to the transmitter manufacturer.

Note 1:

 At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2:

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects and people.

GUIDANCE AND MANUFAC TURER ’S DECL ARATION- ELEC TROMAGNETIC EMISSIONS

The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the ME 
Equipment should assure that it is used in such an environment.

EMISSION TEST

COMPLIANCE

ELEC TROMAGNETIC ENVIRONMENT - GUIDANCE

RF Emissions
CISPR 11

Group 1

The ME Equipment must emit electromagnetic energy in order to perform its intended 
function. Nearby electronic equipment may be affected.

RF Emissions
CISPR 11

Class B

Suitable for use in all establishments, including domestic environments.

GUIDANCE AND MANUFAC TURER ’S DECL ARATION - ELEC TROMAGNETIC IMMUNIT Y

The ME Equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the 
ME Equipment should assure that it is used in such an environment.

IMMUNIT Y TEST

IEC 60601 TEST  LEVEL

COMPLIANCE LEVEL

ELEC TROMAGNETIC ENVIRONMENT - GUIDANCE

Electrostatic discharge 
(ESD) 

IEC 61000-4-2

+8 kV contact

+15 kV air

+8 kV contact

+15 kV air

Floors should be wood, concrete or ceramic 
tile. If floors are covered with synthetic 
material, the relative humidity should be at 
least 30%.

Power frequency (50 / 
60 Hz) magnetic field.
IEC 61000-4-8

30 A/m

30 A/m

Guidance -  
Power frequency magnetic fields should be 
at levels characteristic of typical location in a 
typical hospital environment. 

Portable and mobile RF communications equipment should be used no closer to any part of the ME Equipment, including 
cables, than the recommended separation distance calculated from the equation applicable to the frequency of the 
transmitter.

IMMUNIT Y TEST

IEC 60601 TEST  LEVEL

COMPLIANCE LEVEL

RECOMMENDED SEPARATION DISTANCE

Radiated RF

IEC 61000-4-3

10 V/m

80 MHz to 2.5 GHz

10 V/m

80 MHz to 800 MHz

800 MHz to 2.5 GHz

where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m).

Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site surveya, 
should be less than the compliance level in each 
frequency rangeb.

Interference may occur in the vicinity of equipment 
marked with the following symbol:

Note 1: 

At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 

objects and people.

a) 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, 

AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due 
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ME 
Equipment is used exceeds the applicable RF compliance level above, the ME Equipment should be observed to verify normal operation. If 
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ME Equipment.

b)

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

TEST SPECIFIC ATIONS FOR ENCLOSURE PORT IMMUNIT Y TO RF WIRELESS COMMUNIC ATION EQUIPMENT

TEST 

FREQUENC Y

BAND (A)

(MHZ)

SERVICE (A)

MODUL ATION (B)

MAXIMUM 

POWER (W)

DISTANCE

(M)

IMMUNIT Y TEST 

LEVEL (V/M)

385

380-395

TETRA 400

Pulse 
modulation (b) 
18 Hz

1,8

0,3

27

450

430-470

GMRS 460, 
FRS 460

FM (c) +/- 5 kHz 
deviation 1 kHz 
sine

2

0,3

28

710

704-787

LTE Band 13, 17

Pulse 
modulation (b) 
217 Hz

0,2

0,3

9

745

780

810

800-960

GSM 800/900, 
TETRA 800, 
iDEN 820, 
CDMA 850, LTE 
Band 5

Pulse 
modulation (b)
18 Hz

2

0,3

28

870

930

1 720

1 700-1 990

GSM 1800; 
CDMA 1900; 
GSM 1900; 
DECT; LTE 
Band 1, 3. 4. 35: 
UMTS

Pulse 
modulation (b)
217 Hz

2

0,3

28

1 845

1 970

2 450

2 400-2 570

Bluetooth, 
WLAN, 802.11 
b/g/n, RFID 
2450, LTE 
Band 7

Pulse 
modulation (b)
217 Hz

2

0,3

28

5 240

5 100-5 800

WLAN 802.11 
a/n

Pulse 
modulation (b)
217 Hz

0,2

0,3

9

5 500

5 785

Note:

 If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME 

EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.

(a)

 For some services, only the uplink frequencies are included.

(b)

 The carrier shall be modulated use a 50% duty cycle square wave signal.

(c)

 As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual 

modulation, it would be worst case.

COMPATIBILITY

This sensor is intended for use only with devices containing Masimo technology. 

Each sensor is designed to operate correctly only on the systems from the original 
device manufacturer. Use of this sensor with other devices may result in no or 
improper performance. 

For Compatibility Information Reference: www. Masimo.com

WARRANTY

Masimo warrants to the initial buyer only that these products, when used in accordance with 
the directions provided with the Products by Masimo, will be free of defects in materials and 
workmanship for a period of six (6) months. Single use products are warranted for single 
patient use only. 

THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS 
SOLD BY MASIMO TO BUYER. MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS 
OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF 
MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO’S SOLE OBLIGATION 
AND BUYER’S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO’S 
OPTION, TO REPAIR OR REPLACE THE PRODUCT.

Отзывы: