background image

 

 

 

 

 

Doc. Ref No.: mtt/om/10 

 

                                       

Issue No.:10 

TT7S Series 

 

 

Input Type 

Span Adjustment 

0-2500 ohms 

> 1K ohms (1uA Resolution) 
>200 ohms (5uA Resolution) 

0-5000 ohms 

> 2K ohms (1uA Resolution) 
>400 ohms (5uA Resolution) 

0-75mV 

>30mV (1uA Resolution) 
>10mV (5uA Resolution) 

0-500 mV 

>200mV (1uA Resolution) 
>50mV   (5uA Resolution) 

0-5 V 

>2V (1uA Resolution) 
>500mV (5uA Resolution) 

0-10 V 

>4V (1uA Resolution) 
>1V (5uA Resolution) 

 

S

AFETY AND WARNING

 

 
To  avoid  Electrostatic  Discharge  (ESD)  to  the 
transmitter,  which  may  cause  permanent  damage, 
always  ground  yourself  by  touching  some  ground 
equipment before configuring the transmitter. 
 
Before installation or beginning of any troubleshooting 
procedures  the  power  to  all  equipment  must  be 
switched  off  and  isolated.  Units  suspected  of  being 
faulty  must  be  disconnected  and  removed  first  and 
brought  to  a  properly  equipped  workshop  for  testing 
and repair.  
 
Component  replacement  and  interval  adjustments 
must  be  made  by  a  company  person  only.  Wiring 
must  be  carried  out  by  personnel,  who  have  basic 
electrical knowledge and practical experience.  
 
All  wiring  must  confirm  to  appropriate  standards  of 
good practice and local codes and regulations. Wiring 
must be suitable for voltage, current, and temperature 
rating  of  the  system.  Beware  not  to  over-tighten  the 
terminal screws. 

 

C

ONFIGURATION AND CONNECTION

 

TT7S series transmitters are configurable 
through configuration software “mTRAN". 
Configuration and calibration should be 
done in nonhazardous area. 
Once configuration is done, parameters are 
changed.  
 

Connection details: 

TT7S10-H

 

 

 
 
 
 
 
 
 

Terminal 1, 2& 4:For RTD 3Wire and 
Potentiometer Input 
Terminal 1, 2, 3 & 4: For RTD 4Wire Input 
Terminal 3 & 5: For T/C & Linear Input

 

Terminal 6 & 7:For Power Supply Input

 

 

TT7S10 

 

 
 
 
 
 
 
 
 
 
 

Terminal 1, 2 &4: For RTD 3Wire and 
Potentiometer Input 
Terminal 1, 2, 3 & 4: For RTD 4Wire Input 
Terminal 3 & 4: For T/C & Linear Input

 

Terminal 5 & 6: Power Supply Input 

 

TT7S11 

 

Terminal 7 & 8: Power Supply Input 
Terminal 2, 3, & 4: For RTD Input  
Only, Potentiometer Input 
Terminal 3 & 4: For T/C & Linear Input 
 

 

Table1: Input Sensor  Range

 

Input Type 

Ranges 

-200 to 1000ºC 

-200 to 1200ºC 

-200 to 1370ºC 

-200 to 400ºC 

450 to 1800ºC 

0 to 1750ºC 

0 to 1750ºC 

-200 to 1300ºC 

Pt-100 3  Wire 

-200 to 850ºC 

Pt-100 4  Wire 

(In

 TT7S10 and 

TT7S10-H) 

-200 to 850ºC 

LINEAR 

0-75mV/ 0-500mV 

LINEAR 

(In

 TT7S10 and 

TT7S10-H) 

0-5V/0-10V 

Potentiometer 

0-2500

 

Potentiometer 

(In

 TT7S10 and 

TT7S10-H) 

0-5000

 