3.2.1 The Status Bar
The Status Bar at the top of the display is visible only in
RUN Page. It is divided into five main sections.
1
Time in HH:MM:SS Format
Available in Two Format
1. 24 Hour (default)
2. 12 Hour
This setting is available in Date/Time in Settings Menu
2
Error Code Indicator
This Icon is visible if any On-Board Peripherals like RTC, ADC, DAC, etc.
not working Properly. Refer Section 5.1 on Page for Troubleshooting these
Errors.
The List of Error Code available in this device is given below.
Error Code Description
0
Memory Corrupted or Device Unable to Read/Write
it.
1
RTC Not working Properly
2
Device unable to Read battery Information.
3
Measure Mode Not Working
5
Data Log Memory Corrupt
6
Source Mode Not Working
7
Secondary controller communication error
9
More than one Errors from above list is occurring.
3
USB Connection Status Icon
Icon is visible if USB Charger Adaptor or USB Data Cable is connected to
the Device. Icon is different for both indication & this stated below.
USB Data Cable is connected & Communication with PC
is available.
USB Charger Adaptor is connected.
Battery starts Charging.
4
Battery Charge Percentage Indicator.
Always visible in Run page. Battery % is shown in the center of the icon.
And the icon background is filled with Green, Yellow & Red color if Battery
% is >= 50%, >= 20 %&< 20% respectively.
5
Data Logging Enable Status Indicator.
Icon is visible if Data Logging is enabled and also it will blink when a
Datalog is stored to memory.