6.2.3 Status Display Position Correction
If position correction was enabled in the
Encoder
menu, the
display field appears to the left of the respective position
display (
X
,
Y
,
Z
).
Red, flashing
Machine zero was not yet set or
approached.
The position correction is not active.
Green, constant
Machine zero was set/approached.
The position correction is active.
Note:
For absolute measuring systems,
the machine zero is detected directly
after switching on the PROFILER ST.
Therefore, the position correction is ac-
tive immediately after switching on the
PROFILER ST.
You can select the function for setting the machine zero in the
Origin
menu.
»
Chapter 6.7.3 Origin, page 140
6.2.4 Status Display Reference Signal
If reference signal evaluation was enabled in the
Encoder
menu, the display field
appears to the left of the respective
position display (
X
,
Y
,
Z
).
White
The reference signal evaluation is enabled.
Green
A reference mark was approached.
Содержание PROFILER ST
Страница 2: ......
Страница 3: ...Betriebsanleitung Operating Manual PROFILER ST ...
Страница 72: ...72 Reinigung Betriebsanleitung DE 9 Wartung Der PROFILER ST ist wartungsfrei ...
Страница 98: ...98 Betriebsanleitung DE Notizen Notizen ...
Страница 166: ...166 Cleaning Operating Manual EN 9 Maintenance The PROFILER ST is maintenance free ...
Страница 192: ...192 Operating Manual EN Notes Notes ...
Страница 193: ......
Страница 194: ......
Страница 195: ......