Martini Instruments Mi 105 Скачать руководство пользователя страница 2

WARRANTY

This instrument is warranted against defects in mate-
rials and manufacturing for a period of two years
from the date of purchase. Probe is warranted for 6
months.
If during this period the repair or replacement of parts
is required, where the damage is not due to negli-
gence or erroneous operation by the user, please
return the parts to either distributor or our office and
the repair will be effected free of charge.
Damages due to accidents, misuse, tampering or
lack of prescribed maintenance are not covered.

Notes:

If no valid buffer is detected,
the LCD will show the
“WRNG” message.

For exiting the calibration after the first point
is accepted, press the “ON/OFF” button: the
LCD will show the “OK 1” message for 1
second and then returns to the normal mode.
If using pH 4.01 or 10.01 (or 9.18) buffer
for the first calibration point, the meter will
exit the calibration mode after recognizing
the buffer value and showing the “OK 1”
message for 1 second.

To quit the calibration procedure and return
to the last data, press “ON/OFF” after en-
tering the calibration mode and before the
first point is accepted. The secondary LCD
shows the “ESC” message for 1 second
and the meter returns to the normal mode.

To clear a previous calibration and reset to
the default value, press the “SET” key after
entering the calibration mode and before
the first point is accepted. The secondary
LCD shows the “CLR” message for 1 sec-
ond, the meter resets to the default calibra-
tion and the “CAL” tag on the LCD
disappears.

BATTERY REPLACEMENT

The meter shows the remaining battery per-
centage every time it is switched on. When the
battery level is below 5%, the battery symbol
on the bottom left of the LCD lights up to
indicate a low battery condition.

The meter is also provided with BEPS (Battery
Error Prevention System), which automatically
switches the meter off if the battery level is so
low to cause erroneous readings. It is recom-
mended to replace immediately the battery.

Turn the meter off, re-
move the battery compart-
ment cover from the rear
of the meter and replace
the rundown 9V battery
with a new one.

Install the battery while
paying attention to its po-
larity and reattach the
cover.

SPECIFICATIONS

RANGE (*)

-2.00 to 16.00 pH

-5.0 to 105.0 ºC or 23.0 to 221.0 °F

RESOLUTION

0.01 pH

0.1 °C or 0.1 °F

ACCURACY (@25°C)

±0.02 pH

±0.5°C up to 60°C; ±1°C outside

±1°F up to 140 °F; ±2°F outside

TYPICAL EMC DEVIATION

±0.02 pH

±0.2 °C or ±0.4 °F

TEMPERATURE COMPENSATION

Automatic, from -5 to 80°C

pH CALIBRATION

Automatic, 1 or 2-point

PROBE (included)

MA914BR/1

amplified pH/temperature probe

ENVIRONMENT

0 to 50°C; 100% RH max.

BATTERY

  1 x 9V alkaline (included)

BATTERY LIFE

approx. 500 hours of use

AUTO-OFF

 after 8 minutes of non-use

DIMENSIONS

 200×85×50 mm

WEIGHT

260 g (with battery)

(*) 

The temperature range is limited to 80°C (176°F) if

using the MA914BR/1 probe.

ACCESSORIES

MA914BR/1 Combination amplified pH/

Temperature probe with
BNC & RCA connectors and
1 m cable

M10004B

pH4.01 buffer solution, 20
ml sachet, 25 pcs.

M10006B

pH6.86 buffer solution, 20
ml sachet, 25 pcs.

M10007B

pH7.01 buffer solution, 20
ml sachet, 25 pcs.

M10009B

pH9.18 buffer solution, 20
ml sachet, 25 pcs.

M10010B

pH10.01 buffer solution, 20
ml sachet, 25 pcs.

MA9004

pH4.01 buffer solution, 230
ml bottle

MA9006

pH6.86 buffer solution, 230
ml bottle

MA9007

pH7.01 buffer solution, 230
ml bottle

MA9009

pH9.18 buffer solution, 230
ml bottle

MA9010

pH10.01 buffer solution, 230
ml bottle

MA9015

Electrode storage solution,
230 ml bottle

MA9016

Electrode cleaning solution,
230 ml bottle

M10000B

Electrode rinse solution, 20
ml sachet, 25 pcs.