Martini Instruments Mi 105 Скачать руководство пользователя страница 1

GENERAL OPERATION

The meter is supplied com-
plete with a 9V battery.
Remove the battery com-
partment cover on the
back of the meter and in-
stall the battery while pay-
ing attention to its polarity.

Connect the MA914BR/1
probe to the meter.

Turn the instrument on by
pressing the ON/OFF key.
At start-up, the LCD
shows the percentage of
the remaining battery life
for a few seconds, and
then the current measure-
ment.

Always remove the electrode protective cap
before taking any measurement. If the elec-
trode has been left dry, soak the tip (bottom
2.5 cm) in M10000 rinse solution for a few
minutes.

Immerse the tip (2.5 cm) of the probe into
the sample and stir gently.

Read the pH value when the clock symbol
stops blinking.

The temperature reading can be displayed
in °C or °F unit; to select the desired scale,
press and hold the “ON/OFF” key until “TEMP”
and the current temperature unit are dis-
played on the secondary LCD. Use the “SET”
button to select the unit and then press the
“ON/OFF” key a couple of times to return to
normal mode.

To activate the HOLD function, keep pressed
the “HOLD” key. The measured value will

be frozen on the display and the “HOLD”
message appear on the secondary LCD.

After measurements, switch the meter off
by pressing the “ON/OFF” key. The “OFF”
message appears on the LCD; release the
button.

Notes:

Before taking any  measurement, make sure
that the meter has been calibrated (the “CAL”
tag is displayed on the left lower corner of
the LCD).

Always replace the electrode protective cap
after use with a few drops of storage
solution.

CALIBRATION PROCEDURE

It is recommended to recalibrate the meter at
least once every three weeks.
A) Preparation, Buffer solutions:

1. pH 7.01 / 6.86 (MA9007 / MA9006)
2. pH 4.01 (MA9004) for acidic calibration

(pH < 7) or pH 10.01 / 9.18 (MA9010 /
MA9009
) for alkaline range (pH > 7).

Use two beakers for each buffer solution: one
beaker for rinsing the probe, the other for
calibration. In this way the cross contamination
between solutions is minimized.

B) Procedure:

Select the calibration buffer set: press and
hold the “ON/OFF” key until the LCD shows
“TEMP”. Press again this key and the “BUFF”

USER MANUAL

MI 105

Portable

pH/Temperature Meter

message will appear; then select the de-
sired buffer set with the “SET” key: “7.01
pH BUFF” (for standard solutions: pH 4.01,
7.01, 10.01) or “6.86 pH BUFF” (for NIST
solutions: pH 4.01, 6.86, 9.18). Press the
“ON/OFF” key again to exit.

Remove the protective cap, then immerse
the probe in the first buffer (pH 7.01 or
6.86).

Press and hold the “ON/OFF” key until “CAL”
is displayed on the lower LCD.

Release the button and the
message “7.01 pH USE” (or
“6.86 pH USE” for NIST
buffer set) will be displayed.

The meter automatically recognizes the buffer
value: if a valid buffer is detected, then its
value is displayed on the LCD
together with the “REC” mes-
sage, and the first calibra-
tion point is accepted.

The meter will then require
the second buffer by show-
ing the “pH 4.01 USE” mes-
sage. Immerse the probe in
the second buffer solution
(pH 4.01 or 10.01 or 9.18). If a valid buffer
is recognized, the meter completes the cali-
bration procedure, shows the “OK 2” mes-
sage for a few seconds, then returns to
the normal mode.

ISTMI105R1     06/04

Отзывы: