Martindale Electric EASYPAT 2100 Скачать руководство пользователя страница 2

5

Fig . 2 Display of reading 

(1.00

Ω

)

The instrument can be connected directly to a 230V supply using the 13A plug
fitted on the supply lead or to a 110V supply, using the adaptor supplied. On power
up the EasyPat monitors the supply voltage, adjusts its measurement circuits and
selects the correct test socket accordingly. The display indicates the supply voltage
on which tests will be carried out.

Appliances to be tested should be connected into the appropriate test socket and
should be compatible with the supply voltage to the instrument. If the appliance has an
on/off switch, it should be set to the on position to ensure that the appliance is fully
tested.

If the value for the measured parameter is outside the range of the EasyPat then
the display will show overrange. This is indicated on the display by a single 1. with
no following zeros. Fig. 3 shows the earth continuity display if the measured value
is greater than 1.99

Ω

.

Fig. 3

NB
On the earth bond tests, overrange means above 1.99

Ω

and is a fail. 

On the insulation test it means above 19.9M

Ω

and is a pass.

The wander earth lead is a three-wire lead that forms part of the measuring
circuitry: It is important that the wander earth lead is plugged into the front
panel of the EasyPat 2100 whenever the EasyPat is in use.

6

A red light which is located below the display illuminates if the instrument

 

overheats during a continuity test sequence. All test functions are disabled for as

 

long as this indicator is illuminated and will only be restored after the instrument

 

has cooled and the red indicator has turned off. See Fig. 2.

1.2 Earth / Continuity (8A and 25A)

For this test a voltage of 6V AC is applied between the earth pin of the plug of the

 

appliance being tested and its exposed metal work, via the wander earth lead.

 

The magnitude of the test current is either 8A or 25A selected by the three

 

position rocker switch. The high current applied in this test verifies that the

 

protective earth path will carry the fault current in the event of an insulation

 

breakdown within the appliance.

1.3 Earth / Continuity (

2

00mA)

The test is similar to that described in 1.2 except that a constant current of 

2

00mA

 

DC is applied. This reduced current is required to test IT equipment whose earth

 

path could be damaged by the higher currents.

1.4 Insulation

500V DC is applied between the appliance phase and neutral joined together and

 

earth to ensure that this insulation is at an acceptable level.

1.5 Run Current

The appliance under test is operated at the nominal supply voltage and the

 

current to it is monitored. Care should be taken to ensure that there is no

 

mechanical hazard with this test and that the wander earth lead is disconnected.

1.6 Run Leakage

This test is similar to that described in 1.5 except that the current flow in the earth

 

lead is monitored to check that no potentially hazardous, voltage induced, earth

 

leakage paths are created when the appliance is operating.

1.7 Flash

This is an optional test that may be omitted. On Class I appliances 1.5kV AC is

 

applied between the phase and neutral joined together and earth. The flash gun

 

is not required for Class I appliances but on Class II appliances 3kV AC is applied

 

between the phase and neutral joined together and the tip of the flash gun which

 

is touched onto exposed appliance metalwork. This test is a further verification

 

that functional and supplementary insulation levels have not deteriorated.

Though all tests can be performed in any desired sequence, it is recommended

 

that the Continuity test is always carried out first on Class I appliances and if this

 

results in a fail no further tests should be carried out on the appliance until the

 

fault in its protective earthing has been rectified.

7

For both Class I and Class II appliances, it is recommended that an Insulation test is
successfully carried out before attempting a Run test or a Flash Test.

The results of the tests are displayed on a high contrast 2½ digit LCD which also
indicates the parameter being tested (

Ω

, M

Ω

, mA, A) and also the current supply

voltage which can be either (110 or 230V). See Fig. 2.

1.8 Unpacking & Inspection

Before unpacking the EasyPat, examine the shipping carton for any sign of
damage. Unpack and inspect the EasyPat for any damage. If there is any damage
then consult your distributor immediately.

1.9 Spares & Accessories

Mains Adaptor - 110V plug to 230V Socket MARTL150 *
Wander Earth Lead TL66 *
Flash Gun MARTL166 *
16Amp, FF, 11/4”, HBC Ceramic fuse (Front panel) FUS-FF16 *
IEC Adaptor 230V: 230V 13A plug to IEC320 connector MAREX332 *
Accessory bag, clips to lid of EasyPat and holds earth lead, flash gun etc MARTC151 *

Full range of appliance labels
3 Phase Adaptors

16A, 4 pin

MARTL151

16A, 5 pin

MARTL152

32A, 4 pin

MARTL153

32A, 5 pin

MARTL154

IEC Adaptor 110V: 110V BS4343 plug (yellow) to IEC320 connector MAREX331

*

Included with EasyPat

8

2. SPECIFICATIONS

2.1 Electrical Specification

Supply Voltage: 

110V/230V±10% 50/60Hz

Power Consumption: 

10/220VA (excluding run)

EARTH CONTINUITY TEST (8A & 25A)

Test Voltage:

6V AC nominal with no load

Test Current:

25A AC nom @ 0.1

Ω 

(25A)

8A AC nom @ 0.1

Ω 

(8A)

Display Range:

0 - 1.99

Ω

Accuracy of Indication:

±10% of reading ±2 digits

EARTH CONTINUITY TEST (

2

00mA)

Test Voltage: 
Test Current: 
Range:

130mV DC nominal open circuit

 

2

00mA DC nominal constant current

 

0 - 1.99

Ω

Accuracy of Indication:

±10% of reading ±2 digit

INSULATION TEST

Test Voltage:

500V DC -0% +20% at 0.5M

Ω

Short Circuit Current:    

1.5mA DC nominal

Display Range:

0 - 19.9M

Ω

Accuracy Of Indication:

±5% ±1 digit of reading

RUN CURRENT TEST

Display Range:

0 - 19.9A

Usable Range:

0 - 13.0A (Dependent on mains plug fuse)

Accuracy of Indication:

±10% of reading ±1 digit

RUN LEAKAGE TEST

Display Range:

0 - 19.9mA AC

Accuracy of Indication:

±10% of reading ±1 digit

Отзывы: