Martel BetaProbe TI Скачать руководство пользователя страница 3

This unit was designed to meet the above requirements using:

EN 60079-0:2006 ELECTRICAL APPARATUS FOR 
EXPLOSIVE GAS ATMOSPHERES -- PART 
0: GENERAL REQUIREMENTS (IEC 
60079-0:2004 (MOD))

EN 60079-11:2007 
EXPLOSIVE ATMOSPHERES -- PART 11: 
EQUIPMENT PROTECTION BY INTRINSIC 
SAFETY "I" (IEC 60079-11:2006 (EQV) + 
CORRIGENDUM DEC. 2006 (EQV))

Special Conditions for Safe Use:

 Misuse 

Should the BetaProbe TI/TI+ be overheated or exposed to sudden 
physical shock (i.e. being dropped) it should be examined for any 
damage that may cause a safety concern. If in doubt please return the 
unit for evaluation to Martel Electronics. Please refer to the Customer 
Service Section for contact information.

 WARNING

 

Check for proper operation of the instrument before taking it into a 
hazardous area.  Should the instrument not operate as expected do 
not take the instrument into a hazardous area and return it to the 
manufacturer for inspection.

 WARNING 

The BetaProbe TI/TI+ is intended for installation only in locations 
providing adequate protection against the entry of solid foreign objects 
or water capable of impairing safety. 

 Caution 

To avoid possible damage to calibrator or to equipment  

under test:

If the message changes to “OL” the range limit is exceeded and the  
BetaProbe TI/TI+ must immediately be removed from the heat source to 
prevent damage.

Symbols Used

The following table lists the International Electrical Symbols. Some or all 
of these symbols may be used on the instrument or in this manual.

Symbol Description 

 

Power OFF

 

Power ON

 

Earth ground

 

Risk of Danger. Important information. Refer to manual.

 Battery

 

Hazardous Voltage

 

Conforms to ATEX requirements

 

Conforms to relevant European Union directives.

 

Wheeled bin, conforms to EC directive 2002/96/EC

Hazard Location Information/Approvals

 Ex Hazardous Areas 

An Ex-hazardous area as used in this manual refers to an area made 
hazardous by the potential presence of flammable or explosive vapors. 
These areas are also referred to as hazardous locations.
 

  

 

 

Ex ib IIB T4 Gb (–10°C 

≤ 

Ta 

+50°C) 

  

ITS10ATEX27114X 

 

 

Ex ib IIB T4 Gb  

 

       II 2 G  IECEx ITS10.0049X 

                          0344 

 

Manufactured by Martel Electronics, Inc.,  

 

3 Corporate Park Dr. 

 

Derry, NH, USA

2

3

Содержание BetaProbe TI

Страница 1: ...BetaProbe TI TI High Accuracy Digital Thermometer User Reference Manual...

Страница 2: ...he readings are considered to be stable enough to record the reading Additional features such as min max storage selectable sample rate auto shutoff backlight and damping allow the user to tailor the...

Страница 3: ...tions providing adequate protection against the entry of solid foreign objects or water capable of impairing safety Caution To avoid possible damage to calibrator or to equipment under test If the mes...

Страница 4: ...trend values in either C or F as calculated since power up or since the last reset of the values Press and hold the Min Max key for 2 seconds until CLR is displayed to reset the MIN MAX TREND values 2...

Страница 5: ...r F 0 01 0 1 1 0 or 10 0 When the limit is exceeded it also shows the general direction in which the readings are presently moving The stability level is calculated from a moving sample window of six...

Страница 6: ...or Confirm or cancel the choice to erase logged data ENTER Erase or cancel data erase Damping The Damping function is a running average of readings used filter out noisy temperature sources The choic...

Страница 7: ...ery door and reinstall the hex screw Be sure to use only the approved list of batteries Any substitution will void the safety ratings of this product Battery Manufacturer Type Alkaline AAA Duracell LR...

Страница 8: ...of poor return packaging Out of warranty repairs and recalibration will be subject to specific charges Under no circumstances will Martel Electronics be liable for any device or circumstance beyond th...

Страница 9: ...0219876 Rev D 11 11 www martelcalibrators com e mail sales martelcorp com Tel 603 434 1433 800 821 0023 Fax 603 434 1653 Martel Electronics 3 Corporate Park Drive Derry NH 03038...

Отзывы: