Model T150
Hand-held
Frequency
Calibrator
OPERATING INSTRUCTIONS
www.
.com
[email protected]
1.800.561.8187
Страница 1: ...Model T150 Hand held Frequency Calibrator OPERATING INSTRUCTIONS www com information itm com 1 800 561 8187...
Страница 2: ...en operated in the source mode the user can quickly step the output in 10 of range increments or make smaller changes using the scroll keys Installation Unpacking Remove the Packing List and verify th...
Страница 3: ...k of unit and how the test leads are connected for the specific application Operating Procedure 1 Turn on power and select the desired range by depressing the range select key Your range choices are C...
Страница 4: ...ing from 1 to 100 volts The waveform shape can be square sine or triangular Figure 2 Accuracy The T150 is checked against an NIST Traceable reference before shipment to verify that each range falls wi...
Страница 5: ...min Input Signal Amplitude 1 to 100 Volts Output Waveform 5 V 10 Square Wave Operating Temperature 0 to 50 C Storage Temperature 20 to 60 C Temperature Stability 0 005 of full scale C Power1 9 V Alka...