with the following symbol: 

 

NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people. 

A)

 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and 
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which 
the scale is used exceeds the application RF compliance level above, the scale should be observed to 
verify normal operation. If abnormal performance is observed, additional measures may be necessary, 
such as re-orienting or relocating the scale. 

B)

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 
 
Recommended separation distance between portable and mobile RF communications 
equipment and the M-640. 
 

This scale is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the scale can help prevent electromagnetic interference by 
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) 
and the scale as recommended below, according to the maximum output power of the communications 
equipment. 

 

Rated maximum output 
power of transmitter 
 

W

 

Separation distance according to frequency of transmitter m

 

150 kHz to 80 MHz 

 

d = 1,2

80 MHz to 800 MHz 

 

d = 1,2

P

 

800 MHz to 2,5 GHz 

 

d = 2,3

P

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d 
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is 
the maximum output rating of the transmitter in watts (w) according to the transmitter manufacturer. 
 
NOTE1) At 80 MHz and 800 MHz, the separation distance for the high frequency range applies. 
NOTE2)  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by 
absorption and reflection from structures, objects and people. 

 
 
 
 

 

Page 17 

Содержание M-640

Страница 1: ...USER MANUAL M 640 Please take time to read these instructions before starting to use the scale Version 1 0 07 06...

Страница 2: ...er Supply and Low Battery 6 Operation Basic Functions 7 Operation Advanced Functions 9 Using the Scale with a Printer 13 Using the Scale with Bluetooth 14 EMC Guidance and Manufacturer s Declaration 1...

Страница 3: ...exceeded Product Specification Model M 640 Accuracy Class Class III Capacity Division 300kg x 100g 500kg x 100g 300kg 200g Weight of Scale Approximately 46kg Units of Measure Kg Function Keys ON OFF H...

Страница 4: ...ues specified in the applicable standards Batteries should be kept away from small children If swallowed promptly seek medical assistance If you have any problems contact Marsden your local dealer you...

Страница 5: ...n of manufacturer of medical product including address Electro medical appliance with attachment of type B Dispose of old appliances separately from your household waste This product must be disposed...

Страница 6: ...V 2A into the port on the side of the indicator Installing Replacing the Battery Pack 1 Take out the battery housing 2 The rechargeable battery pack will slide into or out of the housing 3 Check that...

Страница 7: ...the Scale Press the ON OFF button firmly The scale will first test all of the display segments The scale will now show its current software version number The scale will now go into weighing mode and...

Страница 8: ...F button when the scale is turned on The scale will now power down Setting the Scale to Zero If for any reason the scale shows a reading other than 0 0kg it can be reset to zero Press the ZERO key onc...

Страница 9: ...once Allow the patient to get onto the scale After a few seconds the scale will lock on the person s weight When the patient leaves the scale the weight will remain on the display Press HOLD again to...

Страница 10: ...e ZERO key to confirm the height NB There will always be an active flashing digit in the height display unless HOLD is pressed Weigh the patient as normal The display will show the weight height and B...

Страница 11: ...ft digit will flash Enter the preset tare value by using the numeric keys then press the TARE key again to confirm the value Press the ZERO key to return to normal weighing mode To use the Tare functi...

Страница 12: ...period to be edited is selected by pressing the HOLD key E g To input 25 December 2008 8 00 a m Enter the year Press HOLD to confirm and access the date editing field Enter the date E g 12 25 for Dece...

Страница 13: ...sult can be printed Once the person has been weighed and their BMI calculated simply press the PRINT key to produce the following ticket Connecting the TP 2100 Thermal Printer Plug the cable to the pr...

Страница 14: ...HOLD key select ON enable or OFF disable Press the TARE key to confirm the setting Note Disabling the Bluetooth function when not in use will reduce battery power consumption Display the bluEt menu P...

Страница 15: ...that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV co...

Страница 16: ...Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 KHx to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V m Portable and mobile RF communications equipment...

Страница 17: ...ons equipment and the M 640 This scale is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the scale can help prevent ele...

Страница 18: ...he scale and try again If the scale continues to show the error message it indicates a fault with the electronics or wiring High Low Zero Count 1 The scale is above its zero range Please remove any we...

Страница 19: ...Manufacturer s Declaration of Conformity Page 19...

Страница 20: ...09 364293 E mail sales marsdengroup co uk Manufacturing and Distribution Unit 7 Centurion Business Park Coggin Mill Way Rotherham S60 1FB Head Office Unit 1 Genesis Business Park Sheffield Road Rother...

Отзывы: